late shift summary, posted by Andreas, Sergii on 04 Aug 2023, 23:20
|
achievements:
hooked up sample cooling and reached 20K at sample at typical low consumption rate
we archieved targeted resolving power of
~10,000 or 86meV @ 857eV at low temp 20K after indications that heating broadens the elastic line of CoO
we observe sample damage at 30% and |
night shift summary and end of beamtime, posted by Andreas, Martin on 24 Apr 2023, 15:16
|
achievements:
- finished SAXS data recording till 8am on time-resolved skyrmion nucleation and field dependence
- the setup is sensitive to beam transport drifts due to the closed hor-SLIT and the pinhole in FFT. were able to recover beam intensity by simple |
night shift summary and end of beamtime, posted by Andreas, Martin on 24 Apr 2023, 15:16
|
achievements:
- finished SAXS data recording till 8am on time-resolved skyrmion nucleation and field dependence
- the setup is sensitive to beam transport drifts due to the closed hor-SLIT and the pinhole in FFT. were able to recover beam intensity by simple |
night shift summary, posted by Andreas Scherz / Eugenio Paris on 26 Oct 2023, 07:13
|
Night shift summary
Achievements:
- Recovered time0 on SiN (time0 moved >20ps after FEL beam loss in the evening) |
Shift summary, posted by Andreas Scherz on 07 Sep 2023, 23:46
|
Shift summary 07.09.2023
Achievements:
1. found t0 on LSCO x=0.15 at pre-peak O K edge |
SA3 status, posted by Andreas Koch on 08 Feb 2023, 11:14
|
Bunch number control - SA3 FEL imager
As announced today in the XO meeting the SA3 FEL imager has a new feature, see screenshot, encircled:
- In case of danger for the YAG screen the bunches SA3 are set to "0" via the BunchPatternServer (SA3 Main scene). |
Issue, posted by Andreas Koch on 13 Apr 2023, 15:29
|
Trigger issue with SXP_XTD10_IMGES - Exit Slit imager
The external trigger of the Navitar camera was not set, fixed now.
The single pulse trigger of the imager: MCP + camera is tested and correctly working. |
New screens installed in FEL imager SA3, posted by Andreas Koch on 21 Jun 2023, 15:12
|
YAG screen
The damaged screen "YAG2" with Al coating had marks and had been exchanged.
The Imager Watchdog now protects the screen - operational for some months i.e. the number of bunches is shut down to zero if the intensity at the |
Status, posted by Andreas Galler on 11 Feb 2023, 13:55
|
Gun Modulator Problems
Gun Trip at 11:23 and since then the Modulator trips after each restart.
Accelerator Shift Crews and On Call personnel working on fixing the issue. |
Issue, posted by Andreas Galler on 08 Apr 2023, 10:29
|
Changed EPS Interlock condition for SPB_XTD9_IMGPI/SWITCH/SCREEN_OUT
SPB/SFX called DOC this morning for being limitied to two pulses by the EPS (Instrument should have called PRC instead)
After tracing the issue to the SPB_XTD9_IMGPI/SWITCH/SCREEN_OUT switch SPB/SFX decided to take this imager out of the interlock, confirming |
Shift summary, posted by Alexey Zozulya on 24 Feb 2023, 22:57
|
- Measured the speckle patterns using EPIX detector to estimate the beamsize focused down by 1D lens (Row 13).
- Heat load test to NFL #18 with 50 bunches.
- Measured the speckle patterns with 2D focusing lens to estimate the beamsize. wire edge scan? |
Shift summary, posted by Alexey Zozulya on 18 Oct 2023, 02:09
|
Shift summary (p4239)
Achievements:
- HXSS beam at 10.5 keV at 400 uJ pulse energy |
Shift summary, posted by Alexey Zozulya on 23 Oct 2023, 01:10
|
Milestones:
*Finish measurement program on the STO sample at room temperature
*Check mono fpitch and energy rocking curve |
Shift summary, posted by Alexander Pelka on 01 Mar 2023, 16:07
|
Day shift 28.02
Got the beam, aligned it up to the TCC, mirrors in 3 mrad configuration, checked positions for the CRLs. Optimised
beam at the HAPG. Beam stability was fine, energy up to 1,5 mJ at 8.2 keV, and up to 1 mJ at 9.1 keV (used for the spectrometer calibration). |
HED Shift Summary 07.04., posted by Alejandro Laso Garcia on 07 Apr 2023, 23:00
|
Morning - Users from proposal 3455 continued with their measurement program. Beam stable during the measurements.
Afternoon - Changed the setup from 3455 to 3326.
Evening - Beginning of the alignment for users. Beam stable. |
HED Full day shift summary 25.04., posted by Alejandro Laso Garcia on 26 Apr 2023, 07:55
|
Morning shift:
- First thing in the morning ~7 am, the MDL in the SASE 2 topic were down. Even the watchdog which prevented us from moving SA2 attenuators.
DOC quickly resolved the issue (10 min). |
HED Full Day Summary 26.04., posted by Alejandro Laso Garcia on 27 Apr 2023, 07:51
|
Morning shift:
- Tried to align the fiducial fiber in IC1 for the streak cameras, however it was unsuccessful.
- The karabo device for the SOP had to be restarted. Crosschecked that all the timings were still correct. |
Shift summary, posted by Adam Round on 24 Feb 2023, 05:39
|
Staff: AR, KK
X-ray delivery
9.3keV
Optical laser delivery |
Shift end, posted by Adam Round on 22 Mar 2023, 21:49
|
X-ray delivery
9 keV 1 to 165 pulses ~ 300 mj per pulse - stable delivery during the day
Optical laser delivery
ns lasers for illumination only
Achievements/Observations |
Shift summary, posted by Adam Round on 09 Jun 2023, 07:05
|
Staff: DZ, AR, MK
X-ray delivery
12.42 keV 169 pulses for alignment 16 pulses for data collection. stable delivery excluding the extended interuption
Optical |