Message ID: 287
Entry time: 04 Aug 2023, 23:20
|
Author: |
Andreas, Sergii |
Group: |
SCS |
Subject: |
late shift summary |
|
|
achievements:
- hooked up sample cooling and reached 20K at sample at typical low consumption rate
- we archieved targeted resolving power of ~10,000 or 86meV @ 857eV at low temp 20K after indications that heating broadens the elastic line of CoO
- we observe sample damage at 30% and higher transmissions using 350pulses at 1.13MHz, while 20% or less seems fine and elastic line width and intensity remains stable
- we optimise focus condition == 10microns with focus downstream of sample position. KB VFM = 2.04 (beam size vs KB VFM was measured via knife edge scans) dE ~80-85meV
- transfer new sample holder with test sample
plan for tonight and tomorrow morning
- the exit arm may still improve the resolving power, less may be expected from the grating pitch. this can be explored but has lower priority as we reached slightly above E/dE = 10,000 already.
- take RIXS and XAS data on test sample, find optimum settings (damage threshold, camera readout etc)
- late shift will start with 800nm timing.