OPERATION-2023 logbook, Contact: F.Wolff-Fabris, A.Galler, L.Samoylova, H.Sinn
Not logged in
List
|
Message ID:
94
Entry time:
22 Mar 2023, 21:49
Author:
Adam Round
Group:
SPB
Subject:
Shift end
X-ray delivery
9 keV 1 to 165 pulses ~ 300 mj per pulse - stable delivery during the day
Optical laser delivery
ns lasers for illumination only
Achievements/Observations
Focus alignment
found illumination / shield (100 micron) conditions to optimise background on detector
observed that the flourecence signal on the diode was very low and this appears corelated with drift in the focus - after focus optimisation signal was again visible
several runs collected with varying attenuation with the aim to try to corelate flourecence signal on diode with beam intensity
Issues
Too high background from sample at full transmission required exchange of copper shield form 50 to 100 microns
disc detaching from motor spindle (twice) - first time disc blocked gate valve, second bolt and washer are in the gate valve mechanisme space
in both cases detector had to be warmed to allow venting.