Group Selection Page HELP Controls Group Data Analysis Group ITDM WP76 DET Vacuum EEE WP78 FXE SPB MID HED SQS SXP Sample Environment Operation PSPO XO TS Migrated2Zulip
OPERATION-2025 OPERATION-2024 OPERATION-2023 OPERATION-2022 OPERATION-2021 OPERATION-2020 OPERATION-2019 SASE1-2018 SASE1-2017 Experiment data
  OPERATION-2024 logbook, Contact: F.Wolff-Fabris, A.Galler, H.Sinn, Page 3 of 12  Not logged in ELOG logo
ID Datedown Author Group Subject Text Attachments
  184   29 Sep 2024, 00:07 Chan KimSPBShift summary**X-ray delivery**

- 6 keV, ~2.8 mJ, stable delivery

**Optical laser delivery**
  
  183   28 Sep 2024, 23:42 Jörg HallmannMIDShift summarycontinued with the evaluation of the XPCS
analysis packages
investigated the effect of alignment
  
  182   28 Sep 2024, 01:31 Jörg HallmannMIDShift summary


started
with data acquisition for the ptycho experiment
  
  181   27 Sep 2024, 23:13 Chan KimSPBShift summaryX-ray delivery


6 keV, ~3.0 mJ, stable
delivery


Optical
  
  180   27 Sep 2024, 15:17 Jan GruenertPRCStatusBunch pattern SA1 / SA3

SQS
requests to change the bunch pattern.
Currently SA1/SA3 run interleaved
  
  179   27 Sep 2024, 01:51 Jörg HallmannMIDShift summarystarted initial alignment for the ptycho
experiment 
Continued with XPCS measurement using
  
  178   26 Sep 2024, 23:12 Chan KimSPBShift summaryX-ray delivery


6 keV, ~2.7 mJ, stable
delivery


Optical
  
  177   26 Sep 2024, 01:46 Jörg HallmannMIDShift summaryStarted the day with vacuum & detector
issues
real start of the experiments only
  
  176   25 Sep 2024, 23:36 Chan KimSPBShift summaryX-ray delivery


6 keV, ~2.8 mJ, stable
delivery


Optical
  
  175   25 Sep 2024, 16:53 Chan KimSPBIssue600um thick CVD diamond (XTD2 attenuator)
seems damaged

 
 Screenshot_from_2024-09-25_16-50-57.pngScreenshot_from_2024-09-25_16-51-21.png 
  174   25 Sep 2024, 10:12 Raúl Villanueva-GuerrerootherIssueIssue with SA3 Gas attenuator (SA3_XTD10_GATT)
- SOLVED

Dear all,

after the reported issue by SQS
  
  173   25 Sep 2024, 09:47 Jan GruenertPRCIssueIssue with SA3 Gas attenuator (SA3_XTD10_GATT)

9h06:
SQS informs PRC that the Argon line in GATT
is not reacting on pressure change commands,
 JG_2024-09-29_um_12.06.37.png 
  172   21 Sep 2024, 09:50 Peter ZaldenFXESA1 statusThe xray beam drifted significantly on
wednesday and thursday (18. and 19.) but
then was not drifting on friday (20.), see
 2024-09-21-094858_073864177_exflqr51474.png 
  171   16 Sep 2024, 02:21 Wonhyuk JoMIDShift summary- Pump-probe experiments are being conducted
using PuMa and PP laser. 
- The user program is ongoing over
  
  170   15 Sep 2024, 23:12 Raphael de WijnSPBShift summaryX-ray delivery
8.951 keV, 1.6 mJ up to 16 pulses
at 113 kHz
  
  169   15 Sep 2024, 02:16 Wonhyuk JoMIDShift summary- The seeding optimization is ongoing,
which was not successful. Now we have a total
of 170 uJ including SASE and seeding. 
  
  168   14 Sep 2024, 23:13 Raphael de WijnSPBShift summaryX-ray delivery
8.951 keV, 1.2 mJ up to 30pulses at
113 kHz
  
  167   14 Sep 2024, 02:23 Wonhyuk JoMIDShift summary- The seeding optimization result: Total
450 uJ (Seeding of 150 uJ  + SASE of
300 uJ).  BKR will be able to start
  
  166   14 Sep 2024, 01:33 Peter ZaldenFXEStatusFXE received the beam and did not change
accelerator settings.

Beam alignment to FXE was re-established
  
  165   13 Sep 2024, 23:17 Adam RoundSPBShift endOn Shift: TS,
FT, CK, AR, PS

X-ray
  
ELOG V3.1.4-7c3fd00