Entry   Issue, posted by Romain Letrun on 27 Feb 2024, 09:56 8x
The status of the XTD2 attenuators has been checked during the experiment last week and many were found to cause disturbances in the wavefront ranging
from mild to severe.

Attachment 1: No attenuator
Entry   Shift summary, posted by Romain Letrun on 29 Feb 2024, 10:04 
X-ray delivery


10 keV, up to 160 pulses at 1.1MHz, 1.8 mJ


Optical laser
delivery
Entry   Shift summary, posted by Romain Letrun on 01 Mar 2024, 10:54 
X-ray delivery


10 keV, up to 128 pulses with custom bunch pattern, 1.6-2.0 mJ


Optical laser delivery


NA


Achievements/Observations
Entry   Shift summary, posted by Raphael de Wijn on 04 Mar 2024, 06:52 

X-ray delivery


10 keV, up to 128 pulses with custom bunch pattern, 1.6-2.0 mJ


Optical laser delivery


NA


Achievements/Observations
Entry   Issue, posted by Romain Letrun on 04 Mar 2024, 09:42 xtd2_att_watchdog.png
The number of bunches during the last night shift kept being changed down to 24 by the device SA1_XTD2_WATCHDOG/MDL/ATT1_MONITOR020 acting on SA1_XTD2_BUNCHPATTERN/MDL/CONFIGURATOR.
This was not the case during the previous shifts, even though the pulse energy was comparable. Looking at the history of SA1_XTD2_WATCHDOG/MDL/ATT1_MONITOR020,
it looks like this device was started at 16:26:25 on Friday, 1st March.
Entry   Shift summary, posted by Marcin Sikorski on 05 Mar 2024, 07:41 

10 keV beam
Detector: JUNGFRAU in the burst mode (16 memory cells)
Achievements:

multiple sfx data sets (lysozyme)
with 4 and 8 pulses (at 1.1MHz) per exposure taken


Entry   Status, posted by Romain Letrun on 13 Mar 2024, 10:09 7x
We checked the XTD2 attenuators following the intervention from XRO (compar ewith elog:37, before the intervention).

The wavefront shows similar distortions with the 75 and 150 um as earlier, which is not too surprising since the cracks are vertical.

Attachments 1-5: Beam on SPB_XTD9_SCR with XTD2 attenuator CVDs
Entry   Shift summary, posted by Chan Kim on 13 Mar 2024, 23:41 
X-ray delivery


6 keV, 2.6 mJ/pulse, stable delivery


Optical laser delivery


800nm PP laser, stable delivery


Achievements/Observations
Entry   Shift summary, posted by Chan Kim on 15 Mar 2024, 00:16 
X-ray delivery

6 keV, 2.4 mJ/pulse, stable delivery

Optical laser delivery
Entry   Issue, posted by Romain Letrun on 15 Mar 2024, 07:08 Screenshot_from_2024-03-15_07-07-40.png
There are now two watchdogs for the XTD2 attenuator running at the same time


SA1_XTD2_WATCHDOG/MDL/ATT1_MONITOR020
SA1_XTD2_OPTICS_PROTECTION/MDL/ATT1


that are configured differently,
which results in different pulse limits (see attachment).
Entry   Shift summary, posted by Chan Kim on 15 Mar 2024, 23:43 
**X-ray delivery**

- 6 keV, 3.2 mJ/pulse, stable delivery

**Optical laser delivery**
Entry   Shift summary, posted by Raphael de Wijn on 21 Mar 2024, 06:33 
X-ray delivery
- 9.3 keV, >3 mJ/pulse, stable delivery 

Optical laser delivery
Entry   Shift summary, posted by Raphael de Wijn on 22 Mar 2024, 06:35 
X-ray delivery
- 9.3 keV, >3 mJ/pulse, beam down around 30min

Optical laser delivery
Entry   Shift summary, posted by Chan Kim on 28 Mar 2024, 14:54 
X-ray delivery


12.72 keV, 10 Hz, NOT stable delivery, pointing drastically changed in the afternoon, presumably due to the train-on-demand operation
at SASE2


Optical laser delivery
Entry   Issue, posted by Chan Kim on 28 Mar 2024, 14:56 Screenshot-from-2024-03-28-14-26-18_1.png
Pulse energy fluctuation observed after full train operation in SA2 starts.
Entry   Shift summary, posted by Chan Kim on 29 Mar 2024, 22:38 
X-ray delivery
- 12.72 keV, 10Hz, ~1.4 mJ/pulse, stable delivery 

Optical laser delivery
Entry   Issue, posted by Romain Letrun on 30 Mar 2024, 06:59 Screenshot_from_2024-03-30_06-58-42.png
DOOCS panels are again incorrectly reporting the FXE shutter as opened when it is in fact closed. This had been fixed after it was last reported in February
(elog:24), but reappeared now.
Entry   Issue, posted by Chan Kim on 30 Mar 2024, 22:02 Screenshot_from_2024-03-30_21-59-03.pngScreenshot_from_2024-03-30_21-09-11.png
Continuous pointing drift in horizontal direction.

It may trigger a vacuum spike at SPB OH (MKB_PSLIT section) at 21:07 pm.

 
Entry   Shift summary, posted by Chan Kim on 30 Mar 2024, 22:50 
X-ray delivery

- 12.72 keV, 10Hz, ~1.4 mJ/pulse, continuous beam drift in the horizontal direction observed

Optical laser delivery
Entry   Shift end, posted by Adam Round on 31 Mar 2024, 23:20 
X-ray delivery


12.72 keV, 10Hz, ~1.4 mJ/pulse, some beam drift in the horizontal direction observed (probably mirror related)


Optical
laser delivery