  29   22 Feb 2024, 06:13 Peter ZaldenFXEShift summaryBeam conditions:


Used photon energies between
6 and 7 keV, including short undulator scans
Continuously
  
  38   28 Feb 2024, 06:57 Peter ZaldenFXEIssueNot sure if this is the cause of the fringes
reported before, but at 5.6 keV there are
clear cracks visible on the XTD2 Attenuators
 2024-02-28-064435_203161010_exflqr51474.png2024-02-28-064443_551842636_exflqr51474.png2024-02-28-064454_671377179_exflqr51474.png2024-02-28-085524_727294913_exflqr51474.png 
  39   28 Feb 2024, 08:57 Peter ZaldenFXESA1 status5.61 keV setpoint in SA1 was confirmed
to correspond to 5.660 keV (+/- 0.003 keV),
see attached.
 2024-02-28-081608_626485769_exflqr51474.png 
  58   20 Mar 2024, 23:15 Peter ZaldenFXEShift summaryPreparation for user experiment transient
grating, Cris Svetina


Used 7.12 keV from ACC
  
  60   21 Mar 2024, 23:20 Peter ZaldenFXEShift summaryUser experiment transient grating, Cris
Svetina. (LTP)


Used 7.12 keV from ACC
  
  62   22 Mar 2024, 23:13 Peter ZaldenFXEShift summaryUser experiment transient grating, Cris
Svetina. (LTP)


Used 7.11 keV from
  
  63   23 Mar 2024, 23:12 Peter ZaldenFXEShift summaryUser experiment transient grating, Cris
Svetina. (LTP)


Used 7.11 keV from
  
  82   13 Apr 2024, 23:03 Peter ZaldenFXEShift summaryMeasurements to check the impact of bunch
compression on FXE mono transmission were
successful. We could change the peak current
  
  152   23 Aug 2024, 08:48 Peter ZaldenFXEShift summaryUsed 9.4 keV with 94 kHz wed and thu nights.
Received approx. 0.9 mJ with large fluctuations
in pulse energy, however acceptable for the
  
  159   08 Sep 2024, 20:30 Peter ZaldenFXEIssueWe observe a continuous horizontal beam
drift in SA1 after the M1 incident. It seems
not to converge, so one can wonder where
 2024-09-08-194936_755701995_exflqr51474.png 
  161   09 Sep 2024, 23:19 Peter ZaldenFXEIssueFinally, the drift velocity of the pointing
reduced approx a factor of 2 as compared
to yesterday, see attached. Possibly it will
 2024-09-09-231922_609331430_exflqr51474.png 
  166   14 Sep 2024, 01:33 Peter ZaldenFXEStatusFXE received the beam and did not change
accelerator settings.

Beam alignment to FXE was re-established
  
  172   21 Sep 2024, 09:50 Peter ZaldenFXESA1 statusThe xray beam drifted significantly on
wednesday and thursday (18. and 19.) but
then was not drifting on friday (20.), see
 2024-09-21-094858_073864177_exflqr51474.png 
  195   28 Oct 2024, 09:58 Peter ZaldenFXESA1 statusThis is to report some findings during
operation that should be looked into (mostly
to avoid in the future, because both negatively
 2024-10-28-095823_156722903_exflqr37199.png2024-10-28-100227_584317851_exflqr37199.pngScreenshot_2024-11-08_093316.png 
  68   30 Mar 2024, 00:17 Björn Senfftleben (Resolved) Issues with DAQWe had some trouble with the DAQ. It was
not able to start runs properly.

We got the following error messages:
  