113
|
26 May 2024, 23:23 |
Romain Letrun | SPB | Shift summary | X-ray delivery
6.0 keV, up to 202 pulses
at 564 kHz, 1.8-1.9 mJ.
Stable |
|
112
|
26 May 2024, 21:56 |
Thomas Baumann | SQS | Stability Issue | We asked the machine for two color mode
in the morning. Tuning gave good results
at first (1.6mJ second color, 100uJ first |
|
111
|
26 May 2024, 00:23 |
Romain Letrun | SPB | Shift summary | X-ray delivery
6.0 keV, up to 202 pulses
at 564 kHz and 101 at 282 kHz, 0.8-1.6 mJ
Low |
|
110
|
25 May 2024, 02:38 |
Rebecca Boll | SQS | Issue | Today in the afternoon, SASE1 was tuned
(a description of the activities in att 4).
We now realize that this has in some form |
|
109
|
19 May 2024, 15:20 |
Antje Trapp | XRO | HED 6keV after Euler MDL | |
|
108
|
19 May 2024, 14:49 |
Antje Trapp | XRO | M1 - M2 - M3 euler fine tuning geometriy parameters | We fine tuned geometry parameters for linear
effects on X motion (parallel offset) for
all three mirrors. |
|
107
|
17 May 2024, 10:42 |
Dirk Raiser | SQS | Status | Laser status:
After the laser issues listed in https://in.xfel.eu/elog/OPERATION-2024/99 the |
|
106
|
14 May 2024, 11:49 |
Romain Letrun | SPB | Shift summary | X-ray delivery
11.56 keV, 1-197 pulses
@ 564 kHz, 2.6 mJ.
Optical |
|
105
|
14 May 2024, 09:56 |
Andreas Koch | XPD | Issue | EPS issue for Transmissive imager
SA1 (OTRC.2615.T9) w.r.t Operation eLog #96
The motor screen positions w.r.t. |
|
104
|
13 May 2024, 10:36 |
Romain Letrun | SPB | Shift summary | X-ray
delivery
11.56 |
|
103
|
13 May 2024, 08:48 |
Ulrike Boesenberg | MID | Shift summary | Weekend summary:
Overall stable beam delivery with
300bunches @18keV and 750uJ |
|
102
|
12 May 2024, 07:35 |
Romain Letrun | SPB | Shift summary | X-ray
delivery
11.56 keV, 1-352 pulses |
|
101
|
11 May 2024, 07:20 |
Romain Letrun | SPB | Shift summary | X-ray
delivery
11.56 keV, 1-352 pulses |
|
100
|
10 May 2024, 07:06 |
Romain Letrun | SPB | Shift summary | X-ray
delivery
11.56 keV, 1-352 pulses |
|
99
|
10 May 2024, 00:04 |
Rebecca Boll | SQS | Issue | our pp laser experienced timing jumps by
18 ns today. this happened for the first
time at ~7:30 in the morning without us noticing |
|
98
|
09 May 2024, 10:45 |
Tommaso Mazza | SQS | other | we needed SASE1 to be set to 0 pulses briefly
to check alignment in our experiment.
The FXE control hutch was empty. |
|
97
|
09 May 2024, 07:14 |
Romain Letrun | SPB | Shift summary | X-ray
delivery
11.56 keV, 1-352 pulses |
|
96
|
08 May 2024, 13:51 |
Raśl Villanueva-Guerrero | PRC | Issue | Dear all,
Due to a probable issue at the
SA1 Transmissive imager, the "OUT" |
|
95
|
05 May 2024, 00:25 |
Konopkova Zuzana | HED | Experiment summary |
proposal 5700 finished on 04.05.
The experiment used 352 pulses at
4.5 MHz and 18 keV to X-ray heat |
|
94
|
03 May 2024, 23:31 |
Juncheng E | SPB | Shift summary | X-ray delivery
- 9.3 keV, 3.1 mJ, 214 pulses/train