Message ID: 91     Entry time: 01 May 2024, 16:39
Author: Andreas Galler 
Group: PRC 
Subject: Issue 

840 pulses not possible at SXP due to big brother settings for the EPS Power Limit rule. The minimum pulse energy was assumed to be 6 mJ while the actual pulse XGM reading is 550 uJ.
Solution: The minimum assuemd pulse energy had been lowered from 6 mJ to 2 mJ. This shall be reverted on Tuesday next week.