416
|
26 Oct 2023, 07:10 |
Chan Kim | SPB | Shift summary | X-ray delivery
- 7.8keV, 2.4mj, stable delivery, homogeneos intra train energy most of the time.
Optical laser delivery
- ns laser for particle imaging
Achievements/Observations
- AGIPD BG optimized by opening Q2 & Q4 quadrants a few pixels and adding Pb shield on IRU_SLIT4
- NKB focused optimized
- Data (75 nm Ag cube) and gas BG (no sample) measured with two different gases (N2 and He)
Issues
- AGIPD BG near the center (1-2 pixels) were quite strong even with tight sliting, but issue solved by opening two quadrants (Q2 & Q4) and moving CSS about 1-2 pixels (elog:12731)
- Found additional AGIPD BG in Q2M1 module (mid-Q range), but issue solved by adding Pd shield on IRU_SLIT4 (elog:12734)
|
421
|
27 Oct 2023, 03:19 |
Chan Kim | SPB | Issue | We observed auto changes of the number of X-ray pulses last night and tonight.
It is related to 500 um thick Si attenuator in XTD2_ATT.
Even with 2.4 mm thick CVD (XTD2_ATT) IN, the number of pulses @SA1 changes to 28 whenever we insert 500 um thick Si attenuator.
Watchdog triggered this process. It has to be updated.
First row in the atttachment: The number of X-ray pulses @ SA1
Second row in the attachment: Status of 2.4 mm thick CVD in XTD2_ATT
Third row in the attachment: Status of 0.5 mm thick Si in XTD2_ATT |
422
|
27 Oct 2023, 06:35 |
Chan Kim | SPB | Shift summary | X-ray delivery
- 7.8keV, 2.5mj, stable delivery, homogeneos intra train energy most of the time.
Optical laser delivery
- ns laser for particle imaging
Achievements/Observations
- RGA installed and used for quantitative gas BG analysis
- Data (55 nm Ag cube) and gas BG (no sample) measured with two different gases (N2 and He)
- BG roughly 3.5x better with 3 l/min He flow, but only about 2x better with 3 L/min He + N2/CO2 for stable Tayler cone
Issues
- Pulses often changed from 352 to 28 for no apparent reason, discovered later that it was happening when closing silicon 500 XTD2 arm (elog:421)
|
427
|
31 Oct 2023, 23:08 |
Chan Kim | SPB | Shift summary | X-ray delivery
Optical laser delivery
Achievements/Observations
- Vacuum pipe and He pipe for diamond outcoupling installed
- Aligned beam to downstream
- Saw the beam at tower position (Zyla camera)
|
429
|
01 Nov 2023, 22:57 |
Chan Kim | SPB | Shift summary | X-ray delivery
- 19 keV, ~0.9 mJ/pulse, stable delivery
Optical laser delivery
Achievements/Observations
- Finalized allignment
- Got successful runs with impact from sound eventhough often timing misaligned due to jitter
Issues
- High jitter from output of transducer (Ch 2in oscilloscope) is problematic
|
432
|
02 Nov 2023, 23:25 |
Chan Kim | SPB | Shift summary | X-ray delivery
- 19 keV, ~0.9 mJ/pulse, stable delivery
Optical laser delivery
Achievements/Observations
- Tried many possibilities to fix the jitter issue. We covered everything we could think of, but in the end it was not possible as we could not avoid restricting ourselves from the 0.4s delay required for the ultrasound, making us then suffer from the inter-train jitter.
- Returned to operational experiment setup, eventhough suffering from the jitter issue.
Issues
- High jitter from output of transducer (Ch 2in oscilloscope) could not be solved.
- Turbopumps went off, then we realized multiple power sockets were powerless (F7,8,9,10,11). We connected the turbos to another socket to resume operation, after which the previous sockets were also fixed. We must reconnect the turbos to the typical sockets sometime soon, best expectation to be comming Monday.
|
459
|
24 Nov 2023, 20:48 |
Chan Kim | SPB | other | Beam drift in the horizontal direction (XTD9_XGM x position) started at around 10am and continuously drifted until 16:00.
After then, beam has been drifted back to original position within 4 hours. |
460
|
24 Nov 2023, 23:11 |
Chan Kim | SPB | Shift summary | X-ray delivery
- 16.9 keV, up to 30 pulses and 1.1 MHz for alignment, 1 pulse for data collection
- 900 uJ
- Beam pointing and stability issues towards end of shift
Optical laser delivery
Achievements/Observations
- Crystal diffraction data from 4 different samples collected with detector close up two sample to detector position
Issues
- Jungrfau gain visualisation caused some concern - was confirmed with the hlp of DOC to be a known issue that the Medium and low gains are masked in that region leding to dark areas in corrected map
- sample collision with cone and OSA required realignment of the setup
- pointing stability of beam in the end of the day (BKR confirmed that the electron orbit is okay. So, we may need to contact PRC)
|
461
|
25 Nov 2023, 19:56 |
Chan Kim | SPB | other | Beam drift in the horizontal direction (XTD9_XGM).
From ~16:00, X-ray beam started to drift in the horizontal direction. In total, ~0.3 mm drift observed within 2-3 hours.
Attachment 1: Beam X position in XTD2_XGM and XTD9_XGM |
462
|
25 Nov 2023, 23:45 |
Chan Kim | SPB | Shift summary | X-ray delivery
- 16.9 keV, up to 30 pulses and 1.1 MHz for alignment, 1 pulse for data collection
- 800 uJ
- Beam pointing and stability issues towards end of shift (again)
Optical laser delivery
Achievements/Observations
- Crystal diffraction data from several samples collected with detector close up two sample to detector position
Issues
- The counts in the pupil has been dropping over the day and less than yesterday (one issue was a YAG screen in SPB_XTD9_IMGPI)
- pointing stability of beam in the end of the day (similar to yesterday)
|
458
|
24 Nov 2023, 05:07 |
Bjoern Senfftleben | | Sample Camera got stuck (likely at 11pm) and was only noticed at 5am | After going in and reconnecting the camera, we observe att. 1.
According to logs the camera device server was restarted around 11pm, however the camera instance was never reinstantiated after the restart. The Karabo GUI did not indicate that properly, instead showed a frozen image that looked as if the camera works properly.
(see history of camera state in att. 2)
Device Server log shows that after camera connection was lost, the device server restarted?! Was this restart done on purpose?
2023-11-23T23:02:22.499 ERROR karabo.AravisBasler2Camera : SQS_AQS_UPP/CAM/A2A1920_2: Could not synchronize timestamp: GigEVision write_memory timeout
2023-11-23T23:02:27.504 ERROR karabo.AravisCamera : SQS_AQS_UPP/CAM/A2A1920_2: arv_device_get_float_feature_value failed: GigEVision read_memory timeout
2023-11-23T23:02:30.007 WARN karabo.AravisCamera : SQS_AQS_UPP/CAM/A2A1920_2: Control of the camera exflqr50765 is lost
INFO Successfully loaded plugin: "libslsDetectors.so"
INFO Successfully loaded plugin: "libaravisCameras.so"
2023-11-23T23:02:33.773 INFO openMq : Opened TCP connection to broker %s:%d.
2023-11-23T23:02:33.775 INFO openMq : Connection ping enabled (ping interval = %d second).
2023-11-23T23:02:33.778 INFO openMq : Connection connected to broker
2023-11-23T23:02:33.778 INFO karabo.net.JmsConnection : Opened TCP connection to broker tcp://sa3-br-sys-broker-2:7777
2023-11-23T23:02:33.779 INFO karabo.core.DeviceServer : Logfiles are written to: "/scratch/xctrl/karabo/var/log/cppServer/sqs_nqs_vmi_cam/device-server.log"
2023-11-23T23:02:34.785 INFO karabo.xms.SignalSlotable : Instance starts up in topic 'SQS' as 'cppServer/sqs_nqs_vmi_cam' - Karabo 2.16.6
2023-11-23T23:02:34.786 INFO cppServer/sqs_nqs_vmi_cam : DeviceServer starts up with id: cppServer/sqs_nqs_vmi_cam
2023-11-23T23:02:34.786 INFO cppServer/sqs_nqs_vmi_cam : Updated list of devices available: AlarmService,AravisBasler2Camera,AravisBaslerCamera,AravisCamera,AravisPhotonicScienceCamera,CameraImageSource,DataLoggerManager,FileDataLogger,FileLogReader,Gotthard2Control,Gotthard2Receiver,GotthardControl,GotthardReceiver,GuiServerDevice,ImageSource,InfluxDataLogger,InfluxLogReader,JungfrauControl,JungfrauReceiver,PropertyTest
2023-11-23T23:02:34.934 INFO karabo.core.DeviceServer : Sending instance update as new device plugins are available: AravisBasler2Camera,AravisBaslerCamera
2023-11-23T23:02:34.934 INFO cppServer/sqs_nqs_vmi_cam : Keep watching directory: "/scratch/xctrl/karabo/plugins" for Device plugins
2023-11-23T23:02:34.934 INFO cppServer/sqs_nqs_vmi_cam : Starting Karabo DeviceServer (pid: 235604) on host: sqs-rr-sys-con-cams-1, serverId: cppServer/sqs_nqs_vmi_cam, Broker: tcp://sa3-br-sys-broker-2:7777
2023-11-23T23:02:35.073 INFO karabo.core.DeviceServer : Successfully connected to time server 'SQS_RR_UTC/TSYS/TIMESERVER'
==> Status of Taylor Cone unknown for runs 334 - 401
We had to physically unplug and replug the ethernet cable to the camera to get it to work again. |
117
|
31 Mar 2023, 05:10 |
Benjamin, Yi-Ping, Ru-Pan, Andreas | SCS | Late Shift Summary | Achievements
- Optimized BOZ for Co and Fe
- Estabilished Stable liquid jet flow with 30 mM Fe(bpy)3
Problems
- Jet froze several times
- Chamber warmed up earlier in the day. This led to a "dusting" of CNCbl on many surfaces and in the large turbo. We cleaned the turbo with ethanol.
Plan for Night Shift
- Continue measuring stitic XAS with BOZ scheme.
|
114
|
30 Mar 2023, 04:16 |
Benjamin, Robert, Yi-Ping, Ru-Pan | SCS | Shift Summary | Achievements
- Found coarse timing between XFEL and optical laser
Problems
- Did not see a observe fine timing signal from silicon nitride measurement
Plan for night shift
- Continue with liquid jet optimization and static measurements
|
82
|
15 Mar 2023, 17:01 |
Benjamin, Nahid, Raphael, Philippe, Victoria, Timo | SCS | morning shift summary | Achievements
- Found RIXS signal from liquid jet
- Measured static RIXS and PFY-XAS spectrum of Co(cp)(CO)2
Problems
- hRIXS detector has a data latency issue and images were not being recorded in the DAQ. This required ~2-3 hours to solve plus time lost during night shift.
Plan for Next Shift
- find temporal overlap between X-ray and laser
|
40
|
18 Feb 2023, 07:06 |
Benjamin, Manuel, Ricardo | SCS | Night Shift summary | Achievements
- Used scanning of liquid jet to improve S/N
- Measured x-ray PP XAS as a function of the delay between the two X-ray colors
- Scan energy of probe pulse with fixed pump energy
Problems
- No clear signature of ionized states observed in XAS yet
|
32
|
15 Feb 2023, 06:54 |
Benjamin van Kuiken | SCS | Night Shift Summary | Achievements
- Located flat jet and attempted measurement of XAS at several incident energies 1300, 1305, 1315 eV
Problems
- We have difficulty locating the beam on the sample
- We have not found a signal on eith of the diodes in the Chem chamber.
- Knife edges with the viking are daoable but require more extended analysis
- Jet froze due to a leakage from the side of the chip. The ice crusher has been avoided to solve
- The filter in the hRIXS was broken and had to be relplaced.
|
120
|
02 Apr 2023, 04:46 |
Benjamin van Kuiken | SCS | Late Shift Summary | Achievement
- Measured tXAS at the Fe L-edge of Fe(bpy)3 using both delay and energy scans.
Issues
- The cold traps ran empty on the chamber resulting in a pressure spike that interlocked the DSSC
- The rough vacuum of the chem DPS degraded causing TP5 to go into error. This resulted in valve closing and stopping the measurements
- Jet froze twice
|
352
|
22 Sep 2023, 06:51 |
Benjamin van Kuiken | SCS | Shift summary | Achievemnets
- Characterized the X-ray and laser danage on the STO2LNO2
Problems
- Cound not find fine timeing from XAS measurements performed on the sample
- MCP sees 266 nm PP laser
- SiN mesaurement not successful so far. It is possible that there is an issue with the SN membranes on the sample holder
Plan for next shift
- Find fine timing
- conclude damad sutdy of sample.
|
370
|
27 Sep 2023, 07:51 |
Benjamin van Kuiken | SCS | SCS Night Shift Summary | Achievements
- Characterized optical laser spot at sample
Problems
- FEL beam had very unstable intensity and spectrum. As a result no useful data could be collected during the night
- FEL pointing seemed to be stable on all tunnel imagers. The issue was communicated to BKR, and they said that things looked fine from there perspective.
- Large fluctuations in spectral distrubtion at SCS spectrometer were observed. See video at https://syncandshare.xfel.eu/index.php/s/emtwonCsQJJqpz6
- Oth-order beam on SCS pop-in showed large intensity fluctuations with ~50% of the time showing no beam at all
|
396
|
20 Oct 2023, 15:22 |
Benjamin van Kuiken | SCS | Morning Shift Summary | Achievements
- Tested MCPs and APD for XAS
- Both MCPs give a good XAS spectrum with no amplification, but small MCP requires higher bias voltage
- No fluorescence signal observed on the APD
- Measured XAS of both samples
- Measured static RIXS of under-doped sample
Problems
Plan for next shift
- Sample alignment (half clip+diffraction)
- Switch AFS laser to burst mode and find timing
|
|