X-ray delivery
- 7.8keV, 2.5mj, stable delivery, homogeneos intra train energy most of the time.
Optical laser delivery
- ns laser for particle imaging
Achievements/Observations
- RGA installed and used for quantitative gas BG analysis
- Data (55 nm Ag cube) and gas BG (no sample) measured with two different gases (N2 and He)
- BG roughly 3.5x better with 3 l/min He flow, but only about 2x better with 3 L/min He + N2/CO2 for stable Tayler cone
Issues
- Pulses often changed from 352 to 28 for no apparent reason, discovered later that it was happening when closing silicon 500 XTD2 arm (elog:421)
|