Message ID:
460
Entry time:
24 Nov 2023, 23:11
Author:
Chan Kim
Group:
SPB
Subject:
Shift summary
X-ray delivery
16.9 keV, up to 30 pulses and 1.1 MHz for alignment, 1 pulse for data collection
900 uJ
Beam pointing and stability issues towards end of shift
Optical laser delivery
None
Achievements/Observations
Crystal diffraction data from 4 different samples collected with detector close up two sample to detector position
Issues
Jungrfau gain visualisation caused some concern - was confirmed with the hlp of DOC to be a known issue that the Medium and low gains are masked in that region leding to dark areas in corrected map
sample collision with cone and OSA required realignment of the setup
pointing stability of beam in the end of the day (BKR confirmed that the electron orbit is okay. So, we may need to contact PRC)