449
|
18 Nov 2023, 06:42 |
Tokushi Sato | SPB | Shift summary | X-ray delivery
1.3-1.7 mJ, 10.7 keV, 1-100
pulses, stable delivery
Optical |
|
448
|
17 Nov 2023, 06:55 |
Tokushi Sato | SPB | Shift summary | X-ray delivery
1.3-1.7 mJ, 10.7 keV, 1-100
pulses, stable delivery
Optical |
|
447
|
16 Nov 2023, 06:56 |
Tokushi Sato | SPB | Shift summary | X-ray delivery
1.3-1.7 mJ, 10.7 keV, 1-100
pulses, stable delivery
Optical |
|
446
|
15 Nov 2023, 07:10 |
Jayanath Koliyadu | SPB | Shift end | X-ray delivery
1.3-1.7 mJ, 1-30 pulses,
stable delivery
Optical |
|
445
|
13 Nov 2023, 07:16 |
Loic Le Guyader | SCS | shift summary | # Shift Summary
Achievements
- Measured elastic line with |
|
444
|
12 Nov 2023, 07:24 |
Benjamin van Kuiken | SCS | Night shift summary | # Shift Summary
Achievements
- Found RIXS signal from CuO |
|
443
|
10 Nov 2023, 17:39 |
Jan Gruenert | XPD | SA2 status | End of 30 keV study
The machine finished their attemps
for 30 keV tuning while using NEPAL injector |
|
442
|
10 Nov 2023, 14:10 |
Antje Trapp | XRO | Euler MDL bugfix - motor velocities restored after stop | The Euler MDL was updated, so that velocities
are restored to standard values also after
the stop button was hit before motion was |
|
441
|
07 Nov 2023, 15:57 |
Janni Eidam | | KW44 - Leak in Solid Attenuator | There was a leak in the Solid Attenuator
that increased on Saturday morning when SPB
moved the arm number 4 (see last |
|
440
|
05 Nov 2023, 16:14 |
Jan Gruenert | PRC | Status | Finally, we're completely back in operation,
and the performance is recovered in
all beamlines. |
|
439
|
05 Nov 2023, 12:47 |
Jan Gruenert | PRC | Issue | North branch down : another e-beam
magnets problem
RC informs 12h45 that the linac |
|
438
|
05 Nov 2023, 12:18 |
Jan Gruenert | PRC | Issue | Accelerator down 9h23 - 12h
RC / BKR informs (10am), that due
to a quadrupole power supply failure in A21 |
|
437
|
05 Nov 2023, 04:08 |
Angel Rodriguez Fernandez | MID | Shift summary |
stable SASE beam delivery with
1mJ pulse energy
we took data
for different optical laser excitation fluences |
|
436
|
04 Nov 2023, 13:49 |
Jan Gruenert | PRC | Issue | (Temporary) solution for the vacuum
incident in SA1
This morning at 7h18 the vacuum |
|
435
|
04 Nov 2023, 08:15 |
Jan Gruenert | PRC | Issue | Beam down for SASE1/3: SASE1 XTD2
vacuum issue
Vacuum interlock of the SA1 / XTD2 |
|
434
|
04 Nov 2023, 02:29 |
Johannes Moeller | MID | Shift summary |
Stable delivery of SASE at 18 keV.
The intensity dropped slightly from 1mJ to
0.88mJ during the evening, but everything |
|
433
|
03 Nov 2023, 02:08 |
Johannes Moeller | HED | Shift summary |
stable 18 keV SASE delivery
measurement
of siemens star in the morning
alignment
of phase plate for the nano focussing lenses
testing |
|
432
|
02 Nov 2023, 23:25 |
Chan Kim | SPB | Shift summary | X-ray delivery
19 keV, ~0.9 mJ/pulse,
stable delivery
Optical |
|
431
|
02 Nov 2023, 19:37 |
Jan Gruenert | PRC | Status | Typical delivery status during daytime
during the week:
SA1 at 19 keV : 841 uJ +/- |
|
430
|
02 Nov 2023, 02:28 |
Johannes Moeller | MID | Shift summary |
stable delivery of 1mJ SASE at
18keV
NAFO lens characterization
with caustic scan