X-ray delivery
- 1.3-1.7 mJ, 10.7 keV, 1-100 pulses, stable delivery
Optical laser delivery
Achievements/Observations
- Aligned beam trajectory for 10.7 keV direct beam
- Optimized the microscopy table position and references
- Alingment of spectrometer and first and second crystal
Issues
- Intensity flucutations when SA3 changing number of pulses, it is related to automatic kicker adaptation.
- AGIPD preview update speed was slow
|