Group Selection Page HELP DOC Controls Group Data Analysis Group ITDM WP76 DET Vacuum XRO EEE WP78 WP74 FXE SCS SPB MID HED SQS SXP Sample Environment Photon Commissioning Team Operation PSPO XO TS
OPERATION-2024 OPERATION-2023 OPERATION-2022 OPERATION-2021 OPERATION-2020 OPERATION-2019 SASE1-2018 SASE1-2017 Experiment data
  OPERATION-2024 logbook, Contact: F.Wolff-Fabris, A.Galler, H.Sinn, Page 8 of 10  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Author Group Subject Text Attachments
  60   21 Mar 2024, 23:20 Peter ZaldenFXEShift summaryUser experiment transient grating, Cris
Svetina. (LTP)


Used 7.12 keV from ACC
  
  59   21 Mar 2024, 06:33 Raphael de WijnSPBShift summaryX-ray delivery
- 9.3 keV, >3 mJ/pulse, stable
delivery 
  
  58   20 Mar 2024, 23:15 Peter ZaldenFXEShift summaryPreparation for user experiment transient
grating, Cris Svetina


Used 7.12 keV from ACC
  
  57   20 Mar 2024, 23:03 Giuseppe MercurioSCSShift summary
Late
shift summary


coarse timing
installed
laser shutter on the laser breadboard and
  
  56   20 Mar 2024, 12:41 Frederik Wolff-FabrisPRCIssueBKR called PRC around 06:10AM to inform
on vacuum issues at SA1 beamline leading
to a lost beam permission for the North Branch.
  
  55   16 Mar 2024, 12:12 Harald SinnPRCIssue11:15 SASE1/SPB called that they observed
and issue with the SASE1 XGM/XTD2: about
every 35 seconds the power reading jumps
  
  54   15 Mar 2024, 23:43 Chan KimSPBShift summary**X-ray delivery**

- 6 keV, 3.2 mJ/pulse, stable delivery

**Optical laser delivery**
  
  53   15 Mar 2024, 18:02 Tommaso MazzaSQSStability Issueit is not clear why, but the beam position
on BIU2 has changed over time, even more
since we checked last. it is off in both
 Screenshot_from_2024-03-15_18-04-59.pngScreenshot_from_2024-03-15_18-04-39.png 
  52   15 Mar 2024, 07:08 Romain LetrunSPBIssueThere are now two watchdogs for the XTD2
attenuator running at the same time


SA1_XTD2_WATCHDOG/MDL/ATT1_MONITOR020
SA1_XTD2_OPTICS_PROTECTION/MDL/ATT1


that
 Screenshot_from_2024-03-15_07-07-40.png 
  51   15 Mar 2024, 00:16 Chan KimSPBShift summaryX-ray delivery

6 keV, 2.4 mJ/pulse, stable delivery

Optical laser delivery
  
  50   15 Mar 2024, 00:07 Harald SinnPRCIssueSASE1 bunch number was reduced to 10 pulses
for no obvious reason. 

Traced it back to a malfunctioning
 watchdog_error.jpgcracks_in_SA.jpg 
  49   13 Mar 2024, 23:41 Chan KimSPBShift summaryX-ray delivery


6 keV, 2.6 mJ/pulse, stable
delivery


Optical
  
  48   13 Mar 2024, 10:09 Romain LetrunSPBStatusWe checked the XTD2 attenuators following
the intervention from XRO (compar ewith elog:37, before the intervention).
 7x 
  47   05 Mar 2024, 07:41 Marcin SikorskiSPBShift summary
10 keV beam
Detector:
JUNGFRAU in the burst mode (16 memory cells)
Achievements:

multiple
sfx data sets (lysozyme) with 4 and 8 pulses
  
  46   04 Mar 2024, 09:42 Romain LetrunSPBIssueThe number of bunches during the last night
shift kept being changed down to 24 by the
device SA1_XTD2_WATCHDOG/MDL/ATT1_MONITOR020
 xtd2_att_watchdog.png 
  45   04 Mar 2024, 06:52 Raphael de WijnSPBShift summary
X-ray
delivery


10 keV, up to 128 pulses
  
  44   01 Mar 2024, 17:11 Naresh KujalaPRCStatus-Machine status

the cold compressor failure occurred
at ~16.30hr today. And it will take around
  
  43   01 Mar 2024, 10:54 Romain LetrunSPBShift summaryX-ray delivery


10 keV, up to 128 pulses
with custom bunch pattern, 1.6-2.0 mJ


Optical
  
  42   01 Mar 2024, 06:30 Jörg HallmannMIDShift summary
Shift
summary (Feb 29th):
Stable seeding with 1mJ + 400mJ SASE
  
  41   29 Feb 2024, 10:04 Romain LetrunSPBShift summaryX-ray
delivery


10
  
ELOG V3.1.4-7c3fd00