  470  
 | 
30 Jul 2024, 10:38 | 
Antje Trapp |   | Solid attenuator arm 5 limit switch | On 27th of July, it was reported that arm 
5 of the solid attenuator did not stay 
in 'out' position when retracted, | 
 6x   |  | 
  469  
 | 
18 Apr 2024, 23:19 | 
Jan Gruenert | PRC | Issue | 22h55:
 
HED informs that they have issues 
with the HED pulse picker in XTD6. It affects | 
   |  | 
  467  
 | 
26 Jan 2024, 15:24 | 
Antje Trapp | XRO | end of shift summary | Vertical motors of M3 now calibrated in 
first iteration and gemetrical neutral calibrated 
to 0 in motor settings!!! | 
   |  | 
  466  
 | 
25 Jan 2024, 14:57 | 
Antje Trapp | XRO | end of shift summary | It was possible to align to HED in parasitic 
mode to seeding tests
 
 
mirror horizontal position (coating | 
   |  | 
  465  
 | 
24 Jan 2024, 14:00 | 
Antje Trapp | XRO | scanning for HED beam | We found reflection of M3 on M2 popin, 
but could not locate the beam on HED PBLM 
nor HED_XTD6_SDLBM-1 | 
       |  | 
  464  
 | 
28 Nov 2023, 19:36 | 
Jan Gruenert | XPD | Status | Ramp down activities completed 
by XPD. References:
 
 
 
	SASE1
 
	
 
		https://in.xfel.eu/elog/SASE1+technical+commissioning/2204
 
		https://in.xfel.eu/elog/SASE1+technical+commissioning/2205
 
		https://in.xfel.eu/elog/SASE1+technical+commissioning/2206
 
	
 
	
 
	SASE2
 
	
 
		https://in.xfel.eu/elog/SASE2+technical+commissioning/1206
 
		...until...
 
		https://in.xfel.eu/elog/SASE2+technical+commissioning/1209
 
	
 
	
 
	SASE3
 
	
 
		https://in.xfel.eu/elog/SASE3+technical+commissioning/2220
 
		https://in.xfel.eu/elog/SASE3+technical+commissioning/2221
 
	
 
	
 
 
 
The | 
   |  | 
  463  
 | 
26 Nov 2023, 23:26 | 
Adam Round | SPB | Shift summary | X-ray delivery
 
 
 
	16.9 keV, up to 30 pulses 
and 1.1 MHz for alignment, 1 pulse for data | 
   |  | 
  462  
 | 
25 Nov 2023, 23:45 | 
Chan Kim | SPB | Shift summary | X-ray delivery
 
 
 
	16.9 keV, up to 30 pulses 
and 1.1 MHz for alignment, 1 pulse for data | 
   |  | 
  461  
 | 
25 Nov 2023, 19:56 | 
Chan Kim | SPB | other | Beam drift in the horizontal direction 
(XTD9_XGM).
 
 
From ~16:00, X-ray beam started | 
    |  | 
  460  
 | 
24 Nov 2023, 23:11 | 
Chan Kim | SPB | Shift summary | X-ray delivery
 
 
 
	16.9 keV, up to 30 pulses 
and 1.1 MHz for alignment, 1 pulse for data | 
   |  | 
  459  
 | 
24 Nov 2023, 20:48 | 
Chan Kim | SPB | other | Beam drift in the horizontal direction 
(XTD9_XGM x position) started at around 10am 
and continuously drifted until 16:00. | 
     |  | 
  458  
 | 
24 Nov 2023, 05:07 | 
Bjoern Senfftleben |   | Sample Camera got stuck (likely at 11pm) and was only noticed at 5am | After going in and reconnecting the camera, 
we observe att. 1.
 
 
  | 
      |  | 
  457  
 | 
23 Nov 2023, 23:12 | 
Adam Round | SPB | Shift end | Staff: AR, MS, DM, KK, RL, RdW
 
 
X-ray delivery
 
 
 
	16.9 keV, up to 30 pulses | 
   |  | 
  456  
 | 
23 Nov 2023, 22:15 | 
Jörg Hallmann | MID | Shift summary | 
 
progress 
with the Damnit - XPCS analysis test with 
various samples (SiO2, Ludox etc.) | 
   |  | 
  455  
 | 
22 Nov 2023, 23:18 | 
Raphael de Wijn | SPB | Shift summary | X-ray delivery
 
 
 
	16.9 keV, up to 30 pulses 
and 1.1 MHz for alignment, 1 pulse for data | 
   |  | 
  454  
 | 
22 Nov 2023, 22:48 | 
Johannes Moeller | MID | Shift summary | 
 
*Tests 
for online XPCS pipeline ongoing
 
*beam optimization with power slits, | 
   |  | 
  453  
 | 
22 Nov 2023, 10:19 | 
Johannes Moeller | MID | Issue | The Si 0.5mm absorber of XTD1 attenuator 
seems to be damaged. First screenshot is 
with that one inserted. In the second screenshot, | 
     |  | 
  452  
 | 
21 Nov 2023, 23:29 | 
Adam Round | SPB | Shift end | Staff: AR, MS, DM, RL, RdW
 
 
X-ray delivery
 
 
 
	16.9 keV, up to 30 pulses | 
   |  | 
  451  
 | 
20 Nov 2023, 07:12 | 
Jayanath Koliyadu | SPB | Shift end | X-ray delivery
 
 
 
	1.3-1.7 mJ, 10.7 keV, 1-100 
pulses, stable delivery
 
 
 
Optical | 
   |  | 
  450  
 | 
19 Nov 2023, 06:54 | 
Tokushi Sato | SPB | Shift summary | X-ray delivery
 
 
 
	1.3-1.7 mJ, 10.7 keV, 1-100 
pulses, stable delivery
 
 
 
Optical | 
   |   
 |