X-ray delivery
- 16.9 keV, up to 30 pulses and 1.1 MHz for alignment, 1 pulse for data collection
- 700-800 uJ
- stable delivery
Optical laser delivery
Achievements/Observations
- More data collection with Jungfrau at the back of the CSS
- Crystal diffraction data from salicylic acid, ibuprofen and MOF collected with detector close up
Issues
- Clear aberations from our CRLs, so we operate with as few as possible
- Twice during data collection we lost connection to JF4M module 01 requiering reset of receivers.
|