Entry   Shift summary, posted by Tokushi Sato on 11 Apr 2025, 23:14 
X-ray delivery:

20 keV, ~ 400 uJ, 246 pulses, stable delivery

Optical Laser:
Entry   Shift summary, posted by Tokushi Sato on 12 Apr 2025, 23:15 
X-ray delivery:

20 keV, ~ 500 uJ, 246 pulses, stable delivery

Optical Laser:
Entry   Shift summary, posted by Tokushi Sato on 13 Apr 2025, 22:55 
Staff: PV, TS, JK, FK, + users

X-ray delivery:

20 keV, ~ 300-400 uJ, 250 pulses, stable delivery 
Entry   Shift summary, posted by Tokushi Sato on 14 Apr 2025, 23:08 
X-ray delivery:

20 keV,  250-400 uJ, 250 pulses, stable delivery

Optical Laser:
Entry   Shift summary, posted by Tokushi Sato on 25 Apr 2025, 06:48 
X-ray delivery:

10 keV, ~ 850 uJ, 128 pulses, tuning was needed for 1 hour 

Optical Laser:
Entry   Shift summary, posted by Tokushi Sato on 26 Apr 2025, 06:55 
Staff: PV, TS, JK + external collaborators 

X-ray delivery:

10 keV, ~ 800 uJ, 128 pulses, tuning was needed for 10 min. 
Entry   Shift summary, posted by Tokushi Sato on 01 May 2025, 23:35 
X-ray delivery


8.3 / 9.3 keV, 1400 uJ pulse energy, 1 hour down time


Optical laser delivery


NS illumination
355 nm pump laser


Achievements/Observations
Entry   Shift summary, posted by Tokushi Sato on 03 May 2025, 06:47 
**X-ray delivery**
- 8.3 , 1600 uJ pulse energy

**Optical laser delivery**
Entry   Shift summary, posted by Tokushi Sato on 03 May 2025, 23:13 
X-ray delivery


8.0 keV, 1600 uJ pulse energy


Optical laser delivery


NS illumination
355 nm pump laser


Achievements/Observations
Entry   Shift summary, posted by Tokushi Sato on 04 May 2025, 23:14 
X-ray delivery


8.0 keV, 1300 uJ pulse energy by the end, it lowered during the day


Optical laser
delivery
Entry   Shift summary, posted by Tokushi Sato on 05 May 2025, 22:41 
X-ray delivery


8.0 keV, 1300-1600 uJ pulse energy by the end


Optical laser delivery


NS illumination
355 nm pump laser


Achievements/Observations
Entry   Issue, posted by Romain Letrun on 17 Mar 2025, 06:51 Screenshot_from_2025-03-17_06-45-52.png
The status of the FXE shutter in the Public status display DOOCS panel is incorrect. Here appears as opened, when it is actually closed.
Entry   Shift summary, posted by Romain Letrun on 30 Apr 2025, 22:46 
X-ray delivery


9.3 keV, 1800 uJ pulse energy

1 - 202 pulses at 564 kHz


Poor beam stability overall. Lots of intensity fluctuations and rather long time to recover after change of number of pulses in
Entry   Issue, posted by Romain Letrun on 01 May 2025, 07:25 Screenshot_from_2025-05-01_07-21-57.png
SPB pop-in monitor was not dricing to the switch when we retracted it to the "out " position with the MDL this morning. This prevented us to
use more than two pulses. We have moved the motor one step of 300 um up, which activated the switch and relased the interlock. This should however be fixed
so that the "out" position in the MDL works as expected and the motor operates reliably.
Entry   Shift summary, posted by Raphael de Wijn on 05 Feb 2025, 23:42 
X-ray delivery


up to 202 pulses per train (~1 mJ pulse energy), 0.5 MHz.


Optical laser delivery


PP laser aligned to the interaction point


Achievements/Observations
Entry   Shift summary, posted by Raphael de Wijn on 06 Feb 2025, 23:13 
X-ray delivery


up to 202 pulses per train (~1.4 mJ pulse energy), 0.5 MHz.


Optical laser delivery


OPO issue in the late shift
PP laser aligned and timed both at the interaction point and in PAM


Achievements/Observations
Entry   Issue, posted by Raphael de Wijn on 06 Feb 2025, 23:24 Screenshot_from_2025-02-06_07-48-20.png
Damage on FEL imager was also seen by our morning shift around 7:50am.
Entry   Shift summary, posted by Raphael de Wijn on 08 Feb 2025, 23:44 
SPB/SFX 7955 Day 2

**X-ray delivery**
* up to 202 pulses per train (~1.4 mJ pulse energy), 0.5 MHz.
Entry   Shift summary, posted by Raphael de Wijn on 10 Feb 2025, 00:56 
SPB/SFX 7955 Day 3

X-ray delivery



up to 202 pulses per train (~1.4 mJ pulse energy), 0.5 MHz.
Entry   Shift summary, posted by Raphael de Wijn on 12 Mar 2025, 23:14 

SPB/SFX shift summary user experiment 6933 (setup day)

X-ray delivery
-9.6-9.72 keV with Hirex monitoring, 1.4-1.8 mJ/pulse, 352 pulses/train at 1.1 MHz pulse repetition rate, short pulses