Entry   Shift summary, posted by Jörg Hallmann on 08 Dec 2024, 02:49 
successful measurements of elastomer, gelatine & bio samples

issue:
beam loss at about 2:45 due to communication issues with accelerator modules
Entry   Shift summary, posted by Jörg Hallmann on 07 Dec 2024, 02:04 
data collection of bubbles in elastomers (good progress)

issue: beam loss in the tunnel during the day - experts checked the situation and the situation improved during the evening
Entry   Stability Issue, posted by Johannes Moeller on 06 Dec 2024, 07:49 Screenshot_from_2024-12-06_07-48-48.png
And since the 4th there are some instabilities in the vacuum levels at M1 and M2 as well.
Entry   Stability Issue, posted by Johannes Moeller on 06 Dec 2024, 06:37 Screenshot_from_2024-12-06_06-39-34.png
The jumps keep repeating very frequently since yesterday afternoon. The overall drift of the beam seems to accalerate as well.
Entry   Shift summary, posted by Jörg Hallmann on 06 Dec 2024, 02:41 2024-12-06Bubble0003.png

optimization of the timing of the involved components (laser, cameras, FEL)
generation of the first bubbles in water

issue
Entry   Stability Issue, posted by Alexey Zozulya on 05 Dec 2024, 19:47 2024-12-05_19-43-31.png
Today we have observed multiple large disturbances to beam stability. These can be characterised by large jumps in the horizontal beam position, which
at its extreme amounts to a loss of beam from our behalf. M2 mirror feedbacks are able to return the beam to the initial position after ca. 10 minutes.
Beam jumps have been approximately 1hr apart. We do not observe the beam on the XTD6 tunnel imager in this case, but do observe good lasing quality on
Entry   Issue, posted by Jan Gruenert on 05 Dec 2024, 15:58 
additional comment to MESSAGE ID: 217

PRC was informed by MID at 8h50 about a new appeared limiation to about 60 pulses / train (MID operated previously at 120 pulses/train).
It turned out that this limitation came from the protection of the SA2_XTD1_ATT attenuator, in partiicular the Si attenuator.
Entry   Issue, posted by Johannes Moeller on 05 Dec 2024, 09:18 
MID operation limited to 55 bunches by SA2_XTD1_OPTICS_PROTECTION/MDL/ATT1 because of inserted Si 0.5mm. Seems a bit too tight for 1mJ@17keV, since diamond
is not so usefull at this photon energy anymore.
Entry   Issue, posted by Jan Gruenert on 05 Dec 2024, 07:38 JG_2024-12-05_um_07.30.17.png
Accelerator module issue

7h20: BKR/RC inform that there is was issue with the A12 Klystron filament which had caused
a short downtime (6h15-6h22) and was taken offbeam.
Beam was recovered in all beamlines, but the pulse energy came back lower especially in SA1 (before 920uJ, after about 500 uJ) and higher fluctuations.
Entry   Shift summary, posted by Jörg Hallmann on 05 Dec 2024, 00:06 

machine runs at 17 keV with about 1.2 mJ


achievements:
Entry   Status, posted by Jan Gruenert on 04 Dec 2024, 15:34 JG_2024-12-04_um_15.33.09.png
Beam delivery Status well after tuning finished (which went faster than anticipated).
Entry   Shift summary, posted by Tomas Popelar on 04 Dec 2024, 07:12 
X-ray delivery

19 keV, ~ 400  - 450 uJ, 1 - 128 pulses, stable delivery

Optical laser
Entry   Shift summary, posted by Tokushi Sato on 01 Dec 2024, 23:07 
X-ray delivery

19 keV, ~450 uJ, 1 - 128 pulses, stable delivery

Optical laser
Entry   Shift summary, posted by Tokushi Sato on 30 Nov 2024, 23:01 
X-ray delivery

19 keV, ~550 uJ, 1 - 128 pulses, stable delivery

Optical laser
Entry   SA1 status, posted by Andreas Koch on 26 Nov 2024, 09:47 SA1_FEL_imager_-_damaged_spot_at_YAG_-_original_screen_position.pngSA1_FEL_imager_-_damaged_spot_at_YAG_-_new_screen_position.png
Concerning damage of the SA1 FEL imager, Operation-2024 eLog #206

The screen position of the YAG screen - when pressing the screen selection button - has been moved by 1 mm so that the damaged area is out of the
Entry   Status, posted by Jan Gruenert on 24 Nov 2024, 17:23 JG_2024-11-24_um_17.22.58.png
Status

All beamlines in operation.
The attachment ATT#1 shows the beam delivery of this afternoon, with the pulse energies on the left and the number of bunches on the right.
SA2 was not affected, SA1 was essentially without beam from 13h45 to 16h15 (2.5h) and SA3 didn't get enough (386) pulses in about the same period.
Entry   Issue, posted by Jan Gruenert on 24 Nov 2024, 14:09 JG_2024-11-24_um_16.54.02.png
Beckhoff communication loss SA1-EPS (again)

14h00 Same issue as before, but now EEE cannot clear the problem anymore remotely.
An access is required immediately to repair the SA1-EPS loop to regain communication.
EEE-OCD and VAC-OCD are coming in for the access and repair.
Entry   Issue, posted by Jan Gruenert on 24 Nov 2024, 13:47 
Beckhoff communication loss SA1-EPS

12h19: BKR informs PRC about a limitation to 30 bunches in total for the north branch.
The property "safety_CRL_SA1" is lmiting in MPS.
DOC and EEE-OCD and PRC work on the issue.
Entry   Issue, posted by Jan Gruenert on 23 Nov 2024, 15:11 JG_2024-11-23_um_14.33.06.pngJG_2024-11-23_um_14.40.23.pngJG_2024-11-23_um_15.14.33.png
Damage on YAG screen of SA1_XTD2_IMGFEL

A damage spot was detected during the restart of SA1 beam operation at 14h30.

To confirm that it is not something in the beam itself or a damage on another component (graphite filter, attenuators, ...), the scintillator
is moved from YAG to another scintillator.
Entry   Issue, posted by Jan Gruenert on 23 Nov 2024, 10:27 JG_2024-11-23_um_12.06.35.pngJG_2024-11-23_um_12.10.23.pngJG_2024-11-23_um_14.33.06.png
SA1 Beckhoff issue

SA1 blocked to max. 2 bunches and SA3 limited to max 29 bunches
At 9h46, SA1 goes down to single bunch. PRC is called because SA3 (SCS) cannot get the required >300 bunches anymore. (at 9h36 SA3 was limited
to 29 bunches/train)