Group Selection Page HELP DOC Controls Group Data Analysis Group ITDM WP76 DET Vacuum XRO EEE WP78 WP74 FXE SCS SPB MID HED SQS SXP Sample Environment Photon Commissioning Team Operation PSPO XO TS
OPERATION-2025 OPERATION-2024 OPERATION-2023 OPERATION-2022 OPERATION-2021 OPERATION-2020 OPERATION-2019 SASE1-2018 SASE1-2017 Experiment data
  OPERATION-2024 logbook, Contact: F.Wolff-Fabris, A.Galler, H.Sinn  Not logged in ELOG logo
Message ID: 218     Entry time: 05 Dec 2024, 15:58
Author: Jan Gruenert 
Group: PRC 
Subject: Issue 

additional comment to MESSAGE ID: 217

PRC was informed by MID at 8h50 about a new appeared limiation to about 60 pulses / train (MID operated previously at 120 pulses/train).
It turned out that this limitation came from the protection of the SA2_XTD1_ATT attenuator, in partiicular the Si attenuator.
MID had not noticed this before probably due to the fact that after bringing back A12 online, there was a bit higher pulse energy for MID, and MID changed the attenuation settings.

MID could continue working with the full required 120 pulses / train by inserting more diamond attenuators in XTD1 instead of Silicon and using additional attenuators downstream.