38
|
28 Feb 2024, 06:57 |
Peter Zalden | FXE | Issue | Not sure if this is the cause of the fringes
reported before, but at 5.6 keV there are
clear cracks visible on the XTD2 Attenuators |
   |
46
|
04 Mar 2024, 09:42 |
Romain Letrun | SPB | Issue | The number of bunches during the last night
shift kept being changed down to 24 by the
device SA1_XTD2_WATCHDOG/MDL/ATT1_MONITOR020 |
|
50
|
15 Mar 2024, 00:07 |
Harald Sinn | PRC | Issue | SASE1 bunch number was reduced to 10 pulses
for no obvious reason.
Traced it back to a malfunctioning |
 |
52
|
15 Mar 2024, 07:08 |
Romain Letrun | SPB | Issue | There are now two watchdogs for the XTD2
attenuator running at the same time
SA1_XTD2_WATCHDOG/MDL/ATT1_MONITOR020
SA1_XTD2_OPTICS_PROTECTION/MDL/ATT1
that |
|
55
|
16 Mar 2024, 12:12 |
Harald Sinn | PRC | Issue | 11:15 SASE1/SPB called that they observed
and issue with the SASE1 XGM/XTD2: about
every 35 seconds the power reading jumps |
|
56
|
20 Mar 2024, 12:41 |
Frederik Wolff-Fabris | PRC | Issue | BKR called PRC around 06:10AM to inform
on vacuum issues at SA1 beamline leading
to a lost beam permission for the North Branch. |
|
65
|
28 Mar 2024, 14:56 |
Chan Kim | SPB | Issue | Pulse energy fluctuation observed after
full train operation in SA2 starts. |
|
67
|
30 Mar 2024, 06:59 |
Romain Letrun | SPB | Issue | DOOCS panels are again incorrectly reporting
the FXE shutter as opened when it is in fact
closed. This had been fixed after it was |
|
69
|
30 Mar 2024, 22:02 |
Chan Kim | SPB | Issue | Continuous pointing drift in horizontal
direction.
It may trigger a vacuum spike at |
 |
91
|
01 May 2024, 16:39 |
Andreas Galler | PRC | Issue | 840 pulses not possible at SXP due to big
brother settings for the EPS Power Limit
rule. The minimum pulse energy was assumed |
|
96
|
08 May 2024, 13:51 |
Raśl Villanueva-Guerrero | PRC | Issue | Dear all,
Due to a probable issue at the
SA1 Transmissive imager, the "OUT" |
   |
99
|
10 May 2024, 00:04 |
Rebecca Boll | SQS | Issue | our pp laser experienced timing jumps by
18 ns today. this happened for the first
time at ~7:30 in the morning without us noticing |
|
105
|
14 May 2024, 09:56 |
Andreas Koch | XPD | Issue | EPS issue for Transmissive imager
SA1 (OTRC.2615.T9) w.r.t Operation eLog #96
The motor screen positions w.r.t. |
|
110
|
25 May 2024, 02:38 |
Rebecca Boll | SQS | Issue | Today in the afternoon, SASE1 was tuned
(a description of the activities in att 4).
We now realize that this has in some form |
   |
131
|
24 Jul 2024, 10:26 |
Harald Sinn | PRC | Issue | In SASE2 undualtor cell 10 has issues.
Suren K. will do a ZZ at 11:30 to fix it
(if possible). Estimated duration: 1-3 hours. |
|
132
|
24 Jul 2024, 11:35 |
Harald Sinn | PRC | Issue | The control rack of the soft mono at SASE3
has some failure. Leandro from EEE will enter
the XTD10 at 12:00. |
|
134
|
27 Jul 2024, 09:10 |
Harald Sinn | PRC | Issue | SASE1: Solid attenuator att5 is in error.
When it is tried to move it out, the arm
lifts, but does not reach the final position |
|
142
|
10 Aug 2024, 07:19 |
Romain Letrun | SPB | Issue | The SASE1 HIREX interlock is not working
as expected.
At the position 'OUT' (step |
|
159
|
08 Sep 2024, 20:30 |
Peter Zalden | FXE | Issue | We observe a continuous horizontal beam
drift in SA1 after the M1 incident. It seems
not to converge, so one can wonder where |
|
161
|
09 Sep 2024, 23:19 |
Peter Zalden | FXE | Issue | Finally, the drift velocity of the pointing
reduced approx a factor of 2 as compared
to yesterday, see attached. Possibly it will |
|