Group Selection Page HELP Controls Group Data Analysis Group ITDM WP76 DET Vacuum EEE WP78 FXE SPB MID HED SQS SXP Sample Environment Operation PSPO XO TS Migrated2Zulip
OPERATION-2025 OPERATION-2024 OPERATION-2023 OPERATION-2022 OPERATION-2021 OPERATION-2020 OPERATION-2019 SASE1-2018 SASE1-2017 Experiment data
  OPERATION-2023 logbook, Contact: F.Wolff-Fabris, A.Galler, L.Samoylova, H.Sinn, Page 24 of 24  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Author Groupdown Subject Text Attachments
  259   17 Jun 2023, 07:41 Faisal Hammad Mekky Koua Strange pulse behavior upstream XGM- Pulse intensity decreases drastically
at XTD2 across the train

- the visualisation of our beam
 Screenshot_from_2023-06-17_07-40-49.png 
  343   14 Sep 2023, 16:49 Thomas Baumann Summary: Night 13th and Day 14thThe late shift on the 13th fixed remaining
problems with the DLD and the calibration
pipeline.
  
  385   12 Oct 2023, 02:00 Tokushi Sato Shift summary 11.10.2023Staff: JK, CK, KL, AS

X-ray delivery


8.4 keV, up to 260 pulses
  
  413   25 Oct 2023, 02:43 Diogo Filipe Monrroy Vilan e Melo SASE 3 background heavily impacts our focusAttch 1: We were having trouble focusing
at IRU chamber. After multiple attempts the
best we could achieve was 2.9-3.1 um beam.
 Screenshot_from_2023-10-25_02-21-29.pngScreenshot_from_2023-10-25_02-36-16.pngScreenshot_from_2023-10-25_02-37-38.pngScreenshot_from_2023-10-25_02-42-21.png 
  412   25 Oct 2023, 06:02 Diogo Filipe Monrroy Vilan e Melo Shift SummaryStaff: DM, CK, KK

X-ray delivery


7.8keV, 2.4mj, stable delivery,
  
  419   26 Oct 2023, 06:48 Diogo Filipe Monrroy Vilan e Melo Shift summary 25.10.2023Staff: DM, CK, KK, JB, SR (SEC), LW (Uppsala
university), MM (GU)

X-ray delivery
  
  441   07 Nov 2023, 15:57 Janni Eidam KW44 - Leak in Solid AttenuatorThere was a leak in the Solid Attenuator
that increased on Saturday morning when SPB
moved the arm number 4 (see last
 20231106_180901.jpg20231106_175457.jpg20231106_175612.jpg20231106_185102_(1).jpg 
  458   24 Nov 2023, 05:07 Bjoern Senfftleben Sample Camera got stuck (likely at 11pm) and was only noticed at 5amAfter going in and reconnecting the camera,
we observe att. 1.

 
 Screenshot_from_2023-11-24_05-01-47.pngScreenshot_from_2023-11-24_05-09-07.pngScreenshot_from_2023-11-24_05-18-07.png 
  470   30 Jul 2024, 10:38 Antje Trapp Solid attenuator arm 5 limit switchOn 27th of July, it was reported that arm
5 of the solid attenuator did not stay
in 'out' position when retracted,
 6x 
ELOG V3.1.4-7c3fd00