beam: 30 keV / 30 uJ 
reduce to single bunch for insertion 
put FEL imager in 
put DD into beam at X=178 mm 
we found the signal 
centering the DD by minimising the diffrence signal 
Bias voltage = 40 V 
when moving Y there seems to be some mechanical bending of the detector holder. The position reached the negative end of the range (approx. -9 mm). 
When moving out and in again we could find the beam at: 
X= 178.629 (centred) 
Y= -3.48 (close to center) 
see screenshot 
run 29 with DD and IMGFEL 
  
The aperture was opened by moving the slits from 2x2 to 4x4 mm 
switch to 2 bunches 
screenshot 
try to center Y position 
stop run 29 
run 30: approx. 5 min at this position 
change FEL-imager ND filter to 0.9 to avoid saturation 
run31 
  
take IMGFEL out and switch to 50 bunches 
actual X center = 178.44 
move left until we loose the signal and scan x from 176.5 to 179.9 
y signal changes -> mechanical coupling 
run 32 
in order to center the Y position we had to move 2 mm down to -5.53 
run 33: 
move left until we loose the signal and scan x from 176.5 to 179.9 
y signal changes 
Now the center of y seems to be at -2.88 mm 
Image was observed with FEL YAG imager, x movement does not result in a horizontal motion but it is some sort of a diagonal motion 
After some movements the y center is now at -5.23 mm 
After a new try the y center is at -3.37 mm 
Due to the mechanical problem with the Y-position we do the scan with the K-mono chamber X-motor 
----- 
see elogs from Jan and Tuba 
----- 
Sascha made an orbit kick with 150 bunches 
due to some numerical problem the kick is disturbed (see screenshot) 
DD can reproduce the kick trace 
take run 40 
Sascha altered the orbit again 150 bunches 
https://ttfinfo.desy.de/XFELelog/show.jsp?dir=/2022/47/26.11_a&pos=2022-11-26T17:20:00 
(see attached graph) 
take run 41 
set bias from 40 V to 100 V 
run 42 
Bias set to 40 V again 
beam changed to 27 keV 
x-calibration scan from 8.3 to 11.7 (K-mono X) 
run 43 
Bias -> 100 V 
run 44 (as before) 
Bias -> 20 V 
run 45 
Suspicion: beam is touching K-mono setup 
move K-mono chamber to X=4.6 mm and move DDM X to 173.55 
Y again has to be adjusted, is now -3.18 
bias set to 40 V 
scan K-mono X from 3.1 to 6.1 
run 46 
bias runs: see Tuba's elog 
   |