| beam: 30 keV / 30 uJreduce to single bunch for insertion
 put FEL imager in
 put DD into beam at X=178 mm
 we found the signal
 centering the DD by minimising the diffrence signal
 Bias voltage = 40 V when moving Y there seems to be some mechanical bending of the detector holder. The position reached the negative end of the range (approx. -9 mm).When moving out and in again we could find the beam at:
 X= 178.629 (centred)
 Y= -3.48 (close to center)
 see screenshot
 run 29 with DD and IMGFEL
 The aperture was opened by moving the slits from 2x2 to 4x4 mm switch to 2 bunches screenshot try to center Y position stop run 29 run 30: approx. 5 min at this position change FEL-imager ND filter to 0.9 to avoid saturationrun31
   take IMGFEL out and switch to 50 bunchesactual X center = 178.44
 move left until we loose the signal and scan x from 176.5 to 179.9y signal changes -> mechanical coupling
 run 32
 in order to center the Y position we had to move 2 mm down to -5.53 run 33:move left until we loose the signal and scan x from 176.5 to 179.9
 y signal changes
 Now the center of y seems to be at -2.88 mm Image was observed with FEL YAG imager, x movement does not result in a horizontal motion but it is some sort of a diagonal motion After some movements the y center is now at -5.23 mm After a new try the y center is at -3.37 mm Due to the mechanical problem with the Y-position we do the scan with the K-mono chamber X-motor -----see elogs from Jan and Tuba
 ----- Sascha made an orbit kick with 150 bunchesdue to some numerical problem the kick is disturbed (see screenshot)
 DD can reproduce the kick trace
 take run 40
 Sascha altered the orbit again 150 buncheshttps://ttfinfo.desy.de/XFELelog/show.jsp?dir=/2022/47/26.11_a&pos=2022-11-26T17:20:00
 (see attached graph)
 take run 41
 set bias from 40 V to 100 V run 42 Bias set to 40 V again beam changed to 27 keV x-calibration scan from 8.3 to 11.7 (K-mono X)run 43
 Bias -> 100 Vrun 44 (as before)
 Bias -> 20 Vrun 45
 Suspicion: beam is touching K-mono setupmove K-mono chamber to X=4.6 mm and move DDM X to 173.55
 Y again has to be adjusted, is now -3.18
 bias set to 40 V scan K-mono X from 3.1 to 6.1run 46
 bias runs: see Tuba's elog   |