ID Date Author Group Subjectdown Text Attachments
  132   10 Apr 2023, 02:25 Jan GruenertXPDIssueREPAIRED / FIXED

Problem: The imager
SPB_XTD9_IMGPII45 insertion motor lost
  
  135   13 Apr 2023, 15:29 Andreas KochXPDIssueTrigger issue
with SXP_XTD10_IMGES - Exit Slit imager

The external trigger of the Navitar
  
  143   19 Apr 2023, 17:51 Laurent MercadierSCSIssueWe report on an issue regarding the bunch
pattern: The bunch pattern read by SCS_RR_UTC/MDL/BUNCH_DECODER
shows 48 pulses, but we instead use 50 pulses.
 bunch_pattern)issue_2023-04-19_17-37-10.png 
  162   28 Apr 2023, 11:21 Naresh KujalaPRCIssueIssue with SA1_XTD9_PBLM and it is stuck.

Hi All,

After discussion with SPB, they
  
  163   28 Apr 2023, 13:00 Naresh KujalaPRCIssue- Received a call from machine RC (Sergey
Tomin)

- There is an issue with DOOCS
  
  176   05 May 2023, 00:50 Tokushi SatoSPBIssueLaser synchronization is gone, unlocked
from 04.05.2023, 22.47. E-mail was sent to
lbsync-expert@desy.de and c.c. to las-op@xfel.eu.
 Screenshot_from_2023-05-05_00-31-48_1.png 
  204   14 May 2023, 05:33 Tommaso MazzaSQSIssueSA3 GATT has trouble in setting the pressure
at ~0.5mbar. setting any value between 0.5
and 0.6 will have it settle around 0.48.
 Screenshot_from_2023-05-14_08-19-46.png 
  214   31 May 2023, 17:35 Naresh KujalaPRCIssue 

HED reported that they are unable
to go 352 pulses and while changing/dailing
 bunchpatternServer-scene_MDL-macros.PNG 
  225   05 Jun 2023, 06:45 Tommaso MazzaSQSIssuearound 6:15, the lights as the one shown
in the attachment turned on by themselves
in the control hutch. after some minutes,
 1685940280680.jpg 
  231   08 Jun 2023, 10:37 Jan GruenertPRCIssueSASE2 vacuum incident

7h10
: MID informs PRC about a vacuum incident
in the SASE2 tunnel which tripped the beam.
 2023-06-08_SA2-beam-delivery.JPG 
  232   08 Jun 2023, 11:02 Johannes MoellerMIDIssueAddition to previous issue:

Last incident (of overall 4), we
had 0 bunches, moved in XTD1 ATT 0.6 and
 Screenshot_from_2023-06-08_11-03-50.pngScreenshot_from_2023-06-08_11-08-06.png 
  233   08 Jun 2023, 17:05 Jan GruenertPRCIssueSA2_XTD1_ATT solid attenuator
vacuum issue

Info
from VAC (Raul Villanueva) 12h30:
 image.png 
  236   09 Jun 2023, 12:37 Jan GruenertPRCIssueRepair of SA2_XTD1_ATT

Status 12h30:
The current MID users took beam until
  
  241   10 Jun 2023, 18:46 Jan GruenertPRCIssue18h05 : RC informs:

Machine is down, because Cryo-OK
was lost.
  
  256   16 Jun 2023, 06:28 Michaela PetrichPRCIssuestaus of JUN16, 11:30

SA1 and SA3 have been adjusted
to acceptable conditions to be compatible
  
  279   29 Jul 2023, 17:57 Harald SinnPRCIssueThe hutch search of SASE3, SXP became defective
around 5am. 

DESY MPS group replaced a defective
  
  297   10 Aug 2023, 22:43 Raśl Villanueva-GuerreroPRCIssueHere is a capture of the observed issue
with the SPB shutter the past night.

 
 image_2023-08-10_225040511.png 
  302   12 Aug 2023, 07:19 Romain LetrunSPBIssuePotentially damaged 1.2 mm CVD in XTD2
attenuator. This feature was not present
earlier in the week.
 Screenshot_from_2023-08-12_07-10-24.pngScreenshot_from_2023-08-12_07-10-32.png 
  345   15 Sep 2023, 14:54 Chan KimSPBIssueVacuum issue at XTD2 tunnel which might
be triggered by SA1_XTD2_ATT Arm5 (1.2mm
thick CVD diamond)
 Screenshot_from_2023-09-15_14-51-56.pngScreenshot_from_2023-09-15_15-06-49.png 
  346   18 Sep 2023, 19:14 Yifeng JiangFXEIssue
We found that the x-ray pulse energy
dropped a lot when we changed the undulator
from 5.5 keV to 5.45 keV. We called BKR to
 11111.PNG3435.PNG 
ELOG V3.1.4-7c3fd00