Tunnel closed today. Might re-open around 14:00
Note that that we re-convene briefly at 17:00 today in MCR (Mock-up Control Room)
Vacuum Loop:
Status:
- Large number of ion pumps not responding properly (2 in XTD2, 63 in XTD9). There is the suspicion that the parameters on some ion pump controllers might be set incorrectly, such as baud rate or the bus address. The generated error code indicates a bad address.
- Test in the lab requires Ion pump controller and PCL. No setup currently readily available. Check in the tunnel required. Could take place between 1200 and 1300 in XTD2 with a ZZ. (Update: B. Racky thinks she will be done around 1400 and would prefer access after this time)
- Preparations needed to include XGM
EPS Loop:
Status:
- Temperature sensors return null value. Suspected firmware bug has been fixed and EPS loop updated by TF
- EPS to MPS interface requires testing. Should be done tomorrow
Motion Loop:
Status:
- Fixes have been applied to the MC2 lib. Tested for TECHNOSOFT driver. Should now work also for PBLMs . Tests do be done starting Wednesday
Imagers
Status:
- Camera driver for SR imager not working
- Camera driver for FEL driver working for imager in XTD2 but not for XTD9
Tasks for today:
- E. Eidam, V. Bondor, , J. Tolkiehn H. Sinn go through all ion pumps in the Silenzi list
- E. Eidam, T. Freyermuth check pump controller settings in the tunnel
- A. Koch and W. Ehsan work on Photonics Camera driver for FEL imager
Tasks for tomorrow:
- M. Planas, T. Maltezopoulos, N. Jardon, J. Eidam test XGM vacuum integration in XTD2 (not XTD9)
- N. Gerasimova, N. Gerasimova and 1 firmware expert test PBLMs in the tunnel
- T. Freyermuth tests EPS -> MPS interface
Other topics
- The two signals from MCP to EPS shall be called MCP-upstream and MCP-downstream
|