Message ID: 334     Entry time: 14 Jul 2017, 12:44
Author: Martin Dommach 
Location: XTD9 
Severity: Problem 
Subject: Fast Valves in XTD9 

Dear all,

the fast valves in XTD9 were causing some trouble. From time to time they collect some electrical noise what triggers the valve. The pressure reading on the valve controller then shows oscillations of up to 10^-5 mbar, while the pressure around the fast valve gauges in the beamline is in the 10^-9 mbar range.
To operate the beamline in the meantime I have opened and disabled both fast valves. 

This needs further investigation on the service day next wednesday.

Best regards,
Martin