Message ID: 204     Entry time: 03 May 2017, 14:38
Author: Raśl Villanueva-Guerrero 
Location: XTD9 
Severity: Feedback 
Subject: Gas Section Interlock on XTD9 - Shut off valve also operative 

 Dear all,

during the test run today by our colleagues of WP74, the interlock condition inside the XGM module was met (small volume pressure burst up to 10-2 mbar range). Now we can see also the expected cut-off event of the gas injection (see attached pictured). It seems that the reason why there was any feedback from the respective shut-off valve was the solenoid actuator on the respective shut-off valve was set in manual mode.

This means that XTD9 Interlock for the whole gas section has been up and running as expected the whole time.

Best Regards.