Message ID: 59     Entry time: 01 May 2025, 07:25
Author: Romain Letrun 
Group: SPB 
Subject: Issue 

SPB pop-in monitor was not dricing to the switch when we retracted it to the "out " position with the MDL this morning. This prevented us to use more than two pulses. We have moved the motor one step of 300 um up, which activated the switch and relased the interlock. This should however be fixed so that the "out" position in the MDL works as expected and the motor operates reliably.

Attachment 1: Screenshot_from_2025-05-01_07-21-57.png  1.199 MB  Uploaded 01 May 2025, 08:26  | Show | Show all