Message ID: 80     Entry time: 10 Apr 2024, 17:03
Author: Kelin Tasca 
Group: XRO 
Subject: M3 calibration 

- M3 TX moved to 0 in steps of 1 mm and beam on FXE Pop-In corrected with M3 RY (+) and M2 RY

- Started calibration with M3 Rz in steps of 0.5 (+) and corrected with M2 RY and RZ to find a position with a minimum of parasitic motion

- Iterations with M3 TY

- CRL 3 and 4 were inserted (using find target)

- The backlash correction was disabled

- Curve for M3 Rz position with Ry correction done. M2 RY/RZ corrections to move beam up and down to have it on the screen

- Curve for M3 Ty position with Ry correction done.

- Curve for M3 Rx position with Ry correction done

M3 vertical motors calibrated to 0

 

Attachment 1: SA1_M1_calibration.ods  26 kB  Uploaded 10 Apr 2024, 22:16
Attachment 2: M3_at_nominal_2024-04-10_21-16-39.png  157 kB  Uploaded 10 Apr 2024, 22:20  | Show | Hide all | Show all
Attachment 3: M3_calibrated_2024-04-10_21-20-34.png  152 kB  Uploaded 10 Apr 2024, 22:21  | Hide | Hide all | Show all