Group Selection Page HELP DOC Controls Group Data Analysis Group ITDM WP76 DET Vacuum XRO EEE WP78 WP74 FXE SCS SPB MID HED SQS SXP Sample Environment Photon Commissioning Team Operation PSPO XO TS
OPERATION-2024 OPERATION-2023 OPERATION-2022 OPERATION-2021 OPERATION-2020 OPERATION-2019 SASE1-2018 SASE1-2017 Experiment data
  OPERATION-2024 logbook, Contact: F.Wolff-Fabris, A.Galler, H.Sinn  Not logged in ELOG logo
Message ID: 38     Entry time: 28 Feb 2024, 06:57
Author: Peter Zalden 
Group: FXE 
Subject: Issue 

Not sure if this is the cause of the fringes reported before, but at 5.6 keV there are clear cracks visible on the XTD2 Attenuators 75 umm and 150um CVD, see attached. There is also a little bit of damage on the 600 um CVD, but not as strong as the others.

At 5.6 keV we will need to use the thin attenuators. I will mention again in case this causes problems downstream.

Edit 8:50h: We do not observe additional interference effects in the downstream beam profile (see attachement 4)

Attachment 1: 2024-02-28-064435_203161010_exflqr51474.png  451 kB  | Show | Hide all | Show all
Attachment 2: 2024-02-28-064443_551842636_exflqr51474.png  391 kB  | Hide | Hide all | Show all
2024-02-28-064443_551842636_exflqr51474.png
Attachment 3: 2024-02-28-064454_671377179_exflqr51474.png  416 kB  | Hide | Hide all | Show all
2024-02-28-064454_671377179_exflqr51474.png
Attachment 4: 2024-02-28-085524_727294913_exflqr51474.png  101 kB  Uploaded 28 Feb 2024, 08:55  | Hide | Hide all | Show all
2024-02-28-085524_727294913_exflqr51474.png
ELOG V3.1.4-7c3fd00