OPERATION-2024 logbook, Contact: F.Wolff-Fabris, A.Galler, H.Sinn
Not logged in
List
|
Message ID:
26
Entry time:
15 Feb 2024, 09:32
Author:
Romain Letrun
Group:
SPB
Subject:
Shift summary
X-ray delivery
1 to ~332 pulses at 6 keV, 4.5 mJ
Stable delivery
Number of pulses limited by the Photon Beam Power local limit that is set to 16 W. Is this intentional? All other limits are set to 40 W
Achievements/Observations
Aligned NKB trajectory
Achieved NKB focus
Reduced background scattering on AGIPD
Started data collection
Issues
Identified a couple of problems with the new DAQ that we were testing. Controls group informed and investigating