Subject: Issue 

Damage on YAG screen of SA1_XTD2_IMGFEL

A damage spot was detected during the restart of SA1 beam operation at 14h30.

To confirm that it is not something in the beam itself or a damage on another component (graphite filter, attenuators, ...), the scintillator is moved from YAG to another scintillator.

ATT#1: shows the damage spot on the YAG (lower left dark spot, the upper right dark spot is the actual beam). Several attenuators are inserted.
ATT#2: shows the YAG when it is just moved away a bit (while moving to another scintillator, but beam still on YAG): only the beam spot is now seen. Attenuators unchanged.
ATT#3: shows the situation with BN scintillator (and all attenuators removed).

XPD expert will be informed about this new detected damage in order to take care.