Group Selection Page HELP DOC Controls Group Data Analysis Group ITDM WP76 DET Vacuum XRO EEE WP78 WP74 FXE SCS SPB MID HED SQS SXP Sample Environment Photon Commissioning Team Operation PSPO XO TS
OPERATION-2024 OPERATION-2023 OPERATION-2022 OPERATION-2021 OPERATION-2020 OPERATION-2019 SASE1-2018 SASE1-2017 Experiment data
  OPERATION-2024 logbook, Contact: F.Wolff-Fabris, A.Galler, H.Sinn  Not logged in ELOG logo
Message ID: 195     Entry time: 28 Oct 2024, 09:58
Author: Peter Zalden 
Group: FXE 
Subject: SA1 status 

This is to report some findings during operation that should be looked into (mostly to avoid in the future, because both negatively affect data quality of the user expt at FXE):

  • At the moment, the 99:1 mode affects 4 to 5 pulses out of 100 in SA1, see first screenshot that shows in the top right the IPM reading with each blue point denoting the average pulse energy in a train. This was not always the case. After initial setup on thursday it did affect only a single train out of 100.
  • On 27.10., the switching of SA2 between 1 and multiple bunches affected SA1 beam heavily, see the jumps in SA1 pulse energy in the second screenshot shown here:
    • The pulse energy dropped by 15% and
    • the pointing changed vertically by 150 um on the FXE PI (data not shown, but available from karabo trendlines)

Edit 8.11.: Added screenshot (grafana/ctrend) to show that the vertical displacement was permanent (not corrected by the feedback)

Attachment 1: 2024-10-28-095823_156722903_exflqr37199.png  65 kB  | Show | Hide all | Show all
Attachment 2: 2024-10-28-100227_584317851_exflqr37199.png  160 kB  | Hide | Hide all | Show all
2024-10-28-100227_584317851_exflqr37199.png
Attachment 3: Screenshot_2024-11-08_093316.png  131 kB  Uploaded 08 Nov 2024, 09:34  | Show | Hide all | Show all
ELOG V3.1.4-7c3fd00