X-ray delivery
- 7.0 keV, up to 202 pulses at 564 kHz, 1.0 mJ.
Optical laser delivery
- ns laser for jet illumination
Achievements/Observations
- Data collection for user sample through column
Issues
- Pressure spike while rotating nozzle tripped AGIPD
- Temperature of ASICs of Q2M1 module too low and not easy to turn it on (issue solved after changing the bath temperature to -27 degreeC)
- Julabo chiller issue again (In the end, set temperature changed to -29.5 degreeC and continued the experiment)
- Another 2 ASICs in Q2M2 and Q2M3 seem unhappy
|