Message ID:
111
Entry time:
26 May 2024, 00:23
Author:
Romain Letrun
Group:
SPB
Subject:
Shift summary
X-ray delivery
6.0 keV, up to 202 pulses at 564 kHz and 101 at 282 kHz, 0.8-1.6 mJ
Low pulse energy after beam handover. Improved around 10:00.
~1.5 hour downtime due to issue with A5.
Several timing jumps of up to 1.5 ps visible on BAM and confirmed by X-ray/optical laser timing measurement at the instrument.
Optical laser delivery
770 nm, up to 202 pulses at 564 kHz and 101 at 282 kHz.
Stable delivery.
Achievements/Observations
.
Collected time-resolved diffraction from gold and gold-platinum nanorods at multiple laser fluences and over long and short delay scans.
Issues
AGIPD interlocked overnight and again during the morning due to temperature.
Had to perform timing measurement and correction multiple times throughout the day due to timing jumps.
ELOG V3.1.4-7c3fd00