Message ID: 108     Entry time: 19 May 2024, 14:49
Author: Antje Trapp 
Group: XRO 
Subject: M1 - M2 - M3 euler fine tuning geometriy parameters 

We fine tuned geometry parameters for linear effects on X motion (parallel offset) for all three mirrors.

dTz was corrected for M2 and M3 to -6.5mm (same geometry)

dTz was corrected for M1 to +6mm

for M2 we were able to do a parabola scan Y motion to check for lrTz

lrTz was then corrected to -433mm

correction in the same direction for M3 did not yield better results

Final mirror positions at HED - see picture