OPERATION-2023 logbook, Contact: F.Wolff-Fabris, A.Galler, L.Samoylova, H.Sinn
Not logged in
List
|
Message ID:
86
Entry time:
17 Mar 2023, 23:43
Author:
Robert, Zhong, Manuel
Group:
SCS
Subject:
Late shift summary 17.03.2023
Activities:
continued tr-RIXS scans
tried to find ways to mitigate the energy drifts of the mono (see below)
confirmed in user data that the FEL-PPL timing was not changing (see below)
setting up for tr-RIXS at beginning of night shift
Problems:
the mono energy has been drifting, presumably due to warming of the grating.
automatic correction of the elastic line in the RIXS data was not working reliably
we suspected a timing drift, but it seems that the monitor (photodiode + oscilloscope) itself is drifting. That's a new one.
ELOG V3.1.4-7c3fd00