Group Selection Page HELP DOC Controls Group Data Analysis Group ITDM WP76 DET Vacuum XRO EEE WP78 WP74 FXE SCS SPB MID HED SQS SXP Sample Environment Photon Commissioning Team Operation PSPO XO TS
OPERATION-2024 OPERATION-2023 OPERATION-2022 OPERATION-2021 OPERATION-2020 OPERATION-2019 SASE1-2018 SASE1-2017 Experiment data
  OPERATION-2023 logbook, Contact: F.Wolff-Fabris, A.Galler, L.Samoylova, H.Sinn  Not logged in ELOG logo
Message ID: 421     Entry time: 27 Oct 2023, 03:19
Author: Chan Kim 
Group: SPB 
Subject: Issue 

We observed auto changes of the number of X-ray pulses last night and tonight.

It is related to 500 um thick Si attenuator in XTD2_ATT.

Even with 2.4 mm thick CVD (XTD2_ATT) IN, the number of pulses @SA1 changes to 28 whenever we insert 500 um thick Si attenuator.

Watchdog triggered this process. It has to be updated.

 

First row in the atttachment: The number of X-ray pulses @ SA1

Second row in the attachment: Status of 2.4 mm thick CVD in XTD2_ATT

Third row in the attachment: Status of 0.5 mm thick Si in XTD2_ATT

Attachment 1: Screenshot_from_2023-10-27_03-09-25.png  107 kB  Uploaded 27 Oct 2023, 04:19  | Show | Show all
ELOG V3.1.4-7c3fd00