Message ID: 421     Entry time: 27 Oct 2023, 03:19
Author: Chan Kim 
Group: SPB 
Subject: Issue 

We observed auto changes of the number of X-ray pulses last night and tonight.

It is related to 500 um thick Si attenuator in XTD2_ATT.

Even with 2.4 mm thick CVD (XTD2_ATT) IN, the number of pulses @SA1 changes to 28 whenever we insert 500 um thick Si attenuator.

Watchdog triggered this process. It has to be updated.

 

First row in the atttachment: The number of X-ray pulses @ SA1

Second row in the attachment: Status of 2.4 mm thick CVD in XTD2_ATT

Third row in the attachment: Status of 0.5 mm thick Si in XTD2_ATT

Attachment 1: Screenshot_from_2023-10-27_03-09-25.png  107 kB  Uploaded 27 Oct 2023, 04:19  | Show | Show all