X-ray delivery
- 9.3 keV, 1-120 pulses,
- good stabiltiy, lowish flux ~900uJ, but experimental requirements were met
Achievements/Observations
- Finished doule slit measurements at micron beam
- Alignment to direct beam
- Sample position for the sand paper found
- Some sand paper test speckle patterns measured
Issues
- Alignment was a bit painful without the NKB and EHC cameras - switch to ZYLA at EHC made ALAS alignment much easier
- diff pump does not transmit much of direct beam without interference and requries tight power slits (known limitation)
Instrument status
- AGIPD HV off.
- AGIPD gate valve closed and pinned.
- Attenuators set to thickest diamond.
- Vacuum valves closed.
- EHU PSLIT closed around direct beam to approx 1.2 x 1.2 mm
- sample moved up by 10mm so both rows accessible in direct beam
- current 'focus' sample position is approx location of sandpaper from evening measurement (lower row, darkside), 10mm in +x moves out of beam
To do for next shift
- continue direct beam speckle measurements (See entry elog:138)
- quite a bit of scattter on AGIPD from direct beam and especially sandpaper samples.
- transmissions of at least ~1 % level are detector safe (no pixels in AGIPD medium gain)
|