OPERATION-2023 logbook, Contact: F.Wolff-Fabris, A.Galler, L.Samoylova, H.Sinn
Not logged in
List
|
Message ID:
397
Entry time:
20 Oct 2023, 23:10
Author:
Romain Letrun
Group:
SPB
Subject:
Shift summary
X-ray delivery
13.6 keV, ~1.6 mJ and up to 352 pulses at 1.1 MHz.
Beam down at 22:47, not recovered before the end of the shift. Otherwise stable delivery.
Optical laser delivery
800 nm, 50 fs, 1 to 176 pulses at 564 kHz pumping TOPAS at 475 nm.
Intensity modulations through the burst during the second half of the day. Solved by LAS OCD. Otherwise stable delivery.
Achievements/Observations
Remounting of diagnostics stack under the injector.
PP laser and X-ray spatial and timing overlap.
Collected data from noon until beam down towards the end of the shift.
Issues
Discrepancies between delay calculated by OpticalDelay device and motor encoder position appearing from time to time.
Karabo GUI crashed twice when after pressing enter while editing a table.