Shift summary for 27.9.2023
X-ray delivery
- 6.3 keV, 3-4 mJ/pulse, 1-352 pulses at 1.13 MHz
Optical laser delivery
- PP laser, 1030 nm, 226 pulses at 1.1 MHz, delivered from 9:45 on and was stabilizing over the day
- ns laser for rayleigh scattering, stabe delivery
Achievements/Observations
- Aligned MKB trajectory and optimzed background
- Scattering laser aligned with X-rays references and overlaped with particle beam
- Few runs with sucrose while trying to optimize injection
Issues
- PP laser was still drifting in energy and pointing, better stabilized towards the end of the shift
- Difficulties with injection, tried sevral times and did not improve much
- Slope in pulse energy along the pulse train, up to 40% change from first to last pulse
- Significant background AGIPD after correcting beam trajectory
|