288
Entry time:
05 Aug 2023, 08:20
Author:
Romain Letrun
Group:
SPB
Subject:
Shift summary
X-ray delivery
9.3 keV, 1-202 pulses at 1.13 MHz, 1.9-2.1 mJ.
Overall stable delivery.
L-IBFB could not be activated due to missing missing transfer matrix elements. Missing parameters entered by the expert who happened to be on shift this morning.
Optical laser delivery
800 nm, 50 fs, 564 kHz. Overlall stable delivery.
Achievements/Observations
Installation and alignment of fixed target sample.
Coarse timing of PP laser.
Found diffraction of sample.
Issues
Retracted CRL positions in MDL SA1_XTD2_CRL/MDL/CRL are not up-to-date and do not drive the CRLs to the "out" switch.
PP laser OXC synchronisation signal dropped out a couple of times. LbSync and LAS informed.