Achievements:
1. Beam in sample chamber, almost overlapping with ALAS
2. We saw signal on MCP and diode2
3. Sample Motors aligned close to the desired height in chamber
Problems:
1. Slow response of Screens and mirror motors, solved by restarting the Beckhoff of the mirror motors
2. Could not find signal of unfocused beam on SCS Pop-In
3. Acquiring data with the spectrometer for energy calibration does not work
4, Motor STZ of XRD does not allow halfclip of the beam.
5. No Fluorescence signal on the APD
6. In the DAQ we can not execute "assign" or "cofigure" action |