255
Entry time:
15 Jun 2023, 23:08
Author:
Johan Bielecki
Group:
SPB
Subject:
Shift summary
X-ray delivery
12.42 keV, stable delivery in single pulse mode with 1.9 mJ/pulse
Optical laser delivery
532 nm ns-laser for sample pumping
Achievements/Observations
Collected protein crystal diffraction for three different samples with varying laser delay and fluence. LCP viscous jet sample delivery.
Stable CRL focus using tunnel collimation with 3 XTD2 lenses followed by 17 focusing lenses in SPB hutch. 2 um FWHM focus with <1um standard deviation in 'X' position.
3 XTD2 lenses gives a round X-ray beam approx 1 mm in size, such that we are almost overfilling the CRL apertues in SPB.
Issues
Some initial jetting issues, which have now been solved.
ELOG V3.1.4-7c3fd00