Shift summary:
- get the beam with 23keV and 120µJ
- falignment of mirrors m1 and m2 on the pt stripe at 23keV
- irst time lens alignment with CRL1 and CRL2, get beam size of 5µm at sample position for users
- took data for the calibration of the AGIPD detector quadrants with LaB6 and Al targets
- set-up trigger of in-vacuum camera and ns-laser to match x-ray arrival time
- could not move the AGPID detector
Beamline status:
- Lenses insert, DN500 closed, AGIPD HV off, Feedback running with 5 pulses and increased attenuation (2mm Si in XTD6 ATT and 0.8 Si OPT ATT additional).
|