Group Selection Page HELP DOC Controls Group Data Analysis Group ITDM WP76 DET Vacuum XRO EEE WP78 WP74 FXE SCS SPB MID HED SQS SXP Sample Environment Photon Commissioning Team Operation PSPO XO TS
OPERATION-2024 OPERATION-2023 OPERATION-2022 OPERATION-2021 OPERATION-2020 OPERATION-2019 SASE1-2018 SASE1-2017 Experiment data
  OPERATION-2023 logbook, Contact: F.Wolff-Fabris, A.Galler, L.Samoylova, H.Sinn  Not logged in ELOG logo
Message ID: 103     Entry time: 24 Mar 2023, 18:27
Author: Tommaso Mazza 
Group: SQS 
Subject: Stability Issue 

the pulse energy in XTD10 is lowered by ~30% when SASE1 changes photon energy, compare att. 1 and 2.

In addition, the intra-train pulse energy stability gets much lower. Potentially, there is an impact on the pointing as well, which is critical for the pump-probe experiment.

Attachment 1: Screenshot_from_2023-03-24_18-25-31.png  93 kB  | Hide | Hide all
Screenshot_from_2023-03-24_18-25-31.png
Attachment 2: Screenshot_from_2023-03-24_18-26-18.png  13 kB  | Hide | Hide all
Screenshot_from_2023-03-24_18-26-18.png
ELOG V3.1.4-7c3fd00