Message ID: 103     Entry time: 24 Mar 2023, 18:27
Author: Tommaso Mazza 
Group: SQS 
Subject: Stability Issue 

the pulse energy in XTD10 is lowered by ~30% when SASE1 changes photon energy, compare att. 1 and 2.

In addition, the intra-train pulse energy stability gets much lower. Potentially, there is an impact on the pointing as well, which is critical for the pump-probe experiment.

Attachment 1: Screenshot_from_2023-03-24_18-25-31.png  93 kB  | Hide | Hide all | Show all