OPERATION-2023 logbook, Contact: F.Wolff-Fabris, A.Galler, L.Samoylova, H.Sinn
Not logged in
List
|
Message ID:
102
Entry time:
23 Mar 2023, 23:24
Author:
Jayanath Koliyadu
Group:
SPB
Subject:
Shift end
X-ray delivery
1-165 pulses, 1.1 MHz - 28.2 kHz, upto 400 uJ @ 50 pC
Optical laser delivery
ns laser for imaging
Achievements/Observations
Optimized sample setup for data collection
Calibrated photodiode @ 8keV.
ADC signal was saturated with amplified photodiode signal.
Removed the amplifier and calibrated photodiode again.
ASE signal is visible without the amplifier
Optimized ASE signal by optimizing the focus and target position in focus
Checking and optiming the focus everytime chaning the sample disk seems to improve the signal
Collected data at 8 keV and 9 keV for about 1-2 hours