Message ID: 101     Entry time: 23 Mar 2023, 14:31
Author: Romain Letrun 
Group: SPB 
Subject: Status 

After SASE1 HIREX GOTTHARD reconfiguration, we now get a single pulse per frame. This was verified by comparing data from HIREX recorded with 564 kHz and 1.13 MHz X-ray repetition rates. Both intensity and number of peaks is comparable for the two repetition rates.

Attachment 1: 20230323_HIREX_integral2.png  118 kB  | Hide | Hide all | Show all
20230323_HIREX_integral2.png
Attachment 2: 20230323_HIREX_mean_spectrum2.png  39 kB  | Show | Hide all | Show all
Attachment 3: 20230323_HIREX_peaks2.png  104 kB  | Show | Hide all | Show all